RULE §289.233Radiation Control Regulations for Radiation Machines Used in Veterinary Medicine

      (i) Each veterinarian possessing a therapeutic radiation machine, a simulator, or an electronic brachytherapy device shall apply for and receive a certificate of registration from the agency before using the device for veterinary use. A person may energize the device for purposes of installation and acceptance testing before receiving a certificate of registration from the agency.

      (ii) The registrant for veterinary use of a therapeutic radiation machine, a simulator, or an electronic brachytherapy device shall be a veterinarian licensed by the Texas Board of Veterinary Medical Examiners.

      (iii) Operators of the therapeutic radiation machine, simulator, or electronic brachytherapy device shall:

        (I) complete device-specific training with a training program provided by the manufacturer; or

        (II) complete training received that is substantially equivalent to the manufacturer's training program from a licensed medical physicist who is trained to use the device; and

        (III) retain a record of each individual's device specific training by the registrant in accordance with subsection (k) of this section for inspection by the agency.

      (iv) Each facility shall develop a quality assurance program in writing or in an electronic reporting system. The quality assurance program shall be implemented as a method of minimizing deviations from facility procedures.

        (I) The quality assurance program shall include the following topics:

          (-a-) Occupational and public radiation safety during simulation; and

          (-b-) Occupational and public radiation safety during treatment.

        (II) Deviations from the operating and safety procedures shall be investigated and brought to the attention of the veterinarian and the RSO.

        (III) The review of the quality assurance program shall be conducted at intervals not to exceed 14 months.

      (v) Written operating and safety procedures shall be developed by a licensed medical physicist with a specialty in therapeutic radiological physics and shall include any restrictions required for the safe operation of the particular therapeutic radiation device. These procedures shall be available in the control area of the therapeutic radiation device. The operators shall be able to demonstrate familiarity with these procedures. These procedures shall include the following:

        (I) therapeutic radiation machines shall not be left unattended unless secured by a locking device which will prevent unauthorized use (A computerized password system would also constitute a locking device);

        (II) when there is a need to immobilize an animal or port film for radiation therapy, mechanical supporting or restraining devices shall be used;

        (III) no individual, other than the animal, shall be in the treatment room during exposures from therapeutic radiation machines operating above 150 kV;

        (IV) at energies less than or equal to 150 kV, any individual in the treatment room shall be protected by a barrier sufficient to meet the requirements of §289.231(m) of this title;

        (V) radiation dose requirements in accordance with §289.231(m) of this title;

        (VI) personnel monitoring requirements in accordance with §289.231(m) of this title;

        (VII) use of protective devices for simulators in accordance with paragraph (3)(H) of this subsection;

        (VIII) procedures for restriction and alignment of beam for simulators in accordance with paragraph (5)(G) of this subsection.

      (vi) The registrant shall perform radiation surveys and contamination smears before the transfer or disposal of an accelerator operating at or above 10 MeV. Such surveys shall be documented and maintained by the registrant for inspection by the agency in accordance with subsection (k)(2) of this section.

    (B) Therapeutic radiation machines capable of operating at energies below 1 MeV.

      (i) Equipment requirements.

        (I) When the tube is operated at its leakage technique factors, the leakage radiation shall not exceed the values specified at the distance stated for the classification of that radiation machine system shown in the following Table II. The leakage technique factors are the maximum-rated peak tube potential and the maximum-rated continuous tube current for the maximum-rated peak tube potential.

Attached Graphic

        (II) Permanent fixed diaphragms or cones used for limiting the useful beam shall provide the same or a higher degree of protection as required for the tube housing assembly.

        (III) Removable and adjustable beam-limiting devices shall meet the following requirements.

          (-a-) Removable beam-limiting devices shall, for the portion of the useful beam to be blocked by these devices, transmit not more than 1.0 percent of the useful beam at the maximum kVp and maximum treatment filter. This requirement does not apply to auxiliary blocks or materials placed in the x-ray field to shape the useful beam to the individual animal.

          (-b-) Adjustable beam-limiting devices installed before March 1, 1989, shall, for the portion of the x-ray beam to be blocked by these devices, transmit not more than 5.0 percent of the useful beam at the maximum kVp and maximum treatment filter.

          (-c-) Adjustable beam-limiting devices installed after March 1, 1989, shall meet the requirements of subclause (I) of this clause.

        (IV) The filter system shall be so designed that:

          (-a-) the filters cannot be accidentally displaced at any possible tube orientation;

          (-b-) for equipment installed after March 1, 1989, an interlock system prevents irradiation if the proper filter is not in place;

          (-c-) the radiation at 5 cm from the filter insertion slot opening does not exceed 30 roentgens per hour (R/hr) (300 mGy/hr) under any operating conditions; and

        (V) The tube housing assembly shall be so marked that it is possible to determine the location of the focal spot to within 5 millimeters (mm), and such marking shall be readily accessible for use during calibration procedures.

        (VI) Contact therapy tube housing assemblies shall have a removable shield of at least 0.5 mm lead equivalency at 100 kVp that can be positioned over the entire useful beam exit port during periods when the beam is not in use.

        (VII) The timer shall:

          (-a-) have a display provided at the treatment control panel and a pre-set time selector; and

          (-b-) not permit an exposure if set at zero;

        (VIII) The control panel, in addition to the displays required in subclause (VII)(-a-) of this clause, shall have the following:

          (-a-) an indication of whether electrical power is available at the control panel and if activation of the x-ray tube is possible;

          (-b-) an indication of whether x-rays are being produced;

          (-c-) means for indicating x-ray tube potential and current;

          (-d-) means for terminating an exposure at any time;

          (-e-) a locking device that will prevent unauthorized use of the therapeutic radiation system (a computerized password system would also constitute a locking device); and

          (-f-) emergency buttons/switches that shall be clearly labeled as to their functions.

        (IX) There shall be means of determining initially the SSD to within 1 cm.

        (X) An indication of shutter position, if applicable, shall appear at the control panel.

      (ii) Facility Requirements

        (I) Each installation shall be provided with primary and secondary barriers as are necessary to assure compliance with §289.231(m) of this title. All protective barriers shall be fixed except for entrance doors or beam interceptors.

        (II) For therapeutic radiation systems capable of operation above 150 kVp.

          (-a-) The control panel shall be located outside the treatment room or in an enclosed booth inside the room.

          (-b-) Interlocks shall be provided such that all entrance doors shall be closed, including doors to any interior booths, before treatment can be initiated or continued. If the radiation beam is interrupted by any door opening, it shall not be possible to restore the machine to operation without closing the door and reinitiating irradiation by manual action at the control panel.

      (iii) Surveys, Calibrations and Spot Checks.

        (I) Surveys shall be performed as follows.

          (-a-) All new and existing facilities not previously surveyed shall have an initial survey made by a licensed medical physicist with a specialty in therapeutic radiological physics, who shall provide a written report of the survey to the registrant. Additional surveys shall be done after any change in the facility, facility design, or equipment that might Cont'd...

Next Page Previous Page