RULE §289.233Radiation Control Regulations for Radiation Machines Used in Veterinary Medicine

        (XVII) Capabilities shall be provided so that all radiation safety interlocks can be checked for correct operation.

      (ii) Facility and shielding requirements.

        (I) Each installation shall be provided with primary and secondary barriers as are necessary to assure compliance with §289.231(m) of this title.

        (II) All protective barriers shall be fixed except for entrance doors or beam interceptors.

        (III) The console shall be located outside the treatment room and all emergency buttons/switches shall be clearly labeled as to their functions.

        (IV) Treatment room entrances shall be provided with a warning light in a readily observable position near the outside of all access doors to indicate when the useful beam is "on."

        (V) Interlocks shall be provided such that all entrance doors shall be closed before treatment can be initiated or continued. If the radiation beam is interrupted by any door opening, it shall not be possible to restore the machine to operation without closing the door and reinitiating irradiation by manual action at the console.

      (iii) Surveys, calibrations, spot checks, and operational requirements.

        (I) Surveys shall be performed as follows.

          (-a-) All new and existing facilities not previously surveyed shall have an initial survey made by a licensed medical physicist with a specialty in therapeutic radiological physics, who shall provide a written report of the survey to the registrant. In addition, such surveys shall be done after any change in the facility or equipment that might cause a significant increase in radiation hazard.

          (-b-) The survey report shall include the following:

            (-1-) a diagram of the facility that details building structures and the position of the console, therapeutic radiation machine, and associated equipment;

            (-2-) a description of the therapeutic radiation system, including the manufacturer, model and serial number, beam type, and beam energy;

            (-3-) a description of the instrumentation used to determine radiation measurements, including the date and source of the most recent calibration for each instrument used;

            (-4-) conditions under which radiation measurements were taken;

            (-5-) projected annual TEDE in areas adjacent to the therapy room; and

            (-6-) a description of workload, use, and occupancy factors employed in determining the projected annual TEDE.

          (-c-) The registrant shall maintain a copy of the survey report and a copy of the survey report shall be provided to the agency within 30 days of completion of the survey. Records of the survey report shall be maintained in accordance with subsection (k)(2) of this section for inspection by the agency.

          (-d-) The survey report shall include documentation of all instances where the installation is in violation of applicable regulations. Any deficiencies detected during the survey shall be corrected before using the machine.

        (II) Calibrations of therapeutic systems shall be performed as follows.

          (-a-) The calibration of systems subject to this subsection shall be performed in accordance with an established calibration protocol before the system is first used for irradiation of an animal and thereafter at time intervals that do not exceed 12 months and after any change that might significantly alter the calibration, spatial distribution, or other characteristics of the therapy beam. The calibration procedures shall be in writing, or documented in an electronic reporting system, and shall have been developed by a licensed medical physicist with a specialty in therapeutic radiological physics.

          (-b-) The calibration shall be performed by a licensed medical physicist with a specialty in therapeutic radiological physics who is physically present at the facility during the calibration.

          (-c-) The calibration shall include verification that the equipment is operating in compliance with the design specifications concerning the light field, animal positioning lasers, and back-pointer lights with the isocenter when applicable, variation in the axis of rotation for the table, gantry, and collimator system, and beam flatness and symmetry at the specified depth.

          (-d-) Records of calibration measurements shall be maintained according to subsection (k)(2) of this section for inspection by the agency.

        (III) The spot check procedures shall be performed in accordance with established protocol, shall be in writing, or documented in an electronic reporting system, and shall have been developed by a licensed medical physicist with a specialty in therapeutic radiological physics.

          (-a-) The written spot check procedures shall specify the frequency at which tests or measurements are to be performed and the acceptable tolerance for each parameter measured in the spot check when compared to the value for that parameter determined in the calibration.

          (-b-) Records of spot check measurements and any necessary corrective actions shall be maintained by the registrant in accordance with subsection (k) of this section for inspection by the agency.

    (D) General requirements for simulators in radiation therapy.

      (i) A technique chart relevant to the particular radiation machine shall be provided or electronically displayed in the vicinity of the console and used by all operators.

      (ii) Each registrant shall have and implement written operating and safety procedures in accordance with paragraph (2) of this subsection.

      (iii) When utilized, protective devices shall follow the requirements of paragraph (3)(H) of this subsection.

      (iv) The operator's position during the exposure shall follow the requirements of paragraph (3)(K) of this subsection.

      (v) No individuals other than the animal and the operators shall be in the treatment room during operation of the simulator.

  (9) Educational facilities. Facilities conducting training using live animals are held to the requirements of this section except for paragraphs (10) - (12) of this subsection concerning film processing.

  (10) Automatic and manual film processing for veterinary facilities and mobile veterinary services.

    (A) Films shall be developed in accordance with the time-temperature relationships recommended by the film manufacturer. The specified developer temperature for automatic processing and the time-temperature chart for manual processing shall be posted in the processing area. If the registrant determines an alternate time-temperature relationship is more appropriate for a specific facility, that time-temperature relationship shall be documented and posted.

    (B) Chemicals shall be replaced according to the chemical manufacturer's or supplier's recommendations or at an interval not to exceed three months.

    (C) Darkroom light leak tests shall be performed at intervals not to exceed six months.

    (D) Lighting in the film processing/loading area shall be maintained with the filter, bulb wattage, and distances recommended by the film manufacturer for that film emulsion or with products that provide an equivalent level of protection against fogging.

    (E) Corrections or repairs of the light leaks or other deficiencies in subparagraphs (B) - (D) of this paragraph shall be initiated within 72 hours of discovery and completed no longer than 15 days from detection of the deficiency unless a longer time is authorized by the agency.

    (F) Documentation of the items in subparagraphs (B), (C), and (E) of this paragraph shall be maintained at the site where performed and shall include the date and initials of the individual completing these items. These records shall be made and maintained in accordance with subsection (k)(2) of this section for inspection by the agency.

  (11) Alternative processing systems. Users of daylight processing systems, laser processors, self-processing film units, or other alternative processing systems shall follow manufacturer's recommendations for image processing. Documentation that the registrant is following manufacturer's recommendations shall include the date and initials of the individual completing the document and shall be made and maintained at the site where performed in accordance with subsection (k)(2) of this section for inspection by the agency.

  (12) Digital imaging acquisition systems.

    (A) Users of digital imaging acquisition systems shall follow quality assurance/quality control (QA/QC) protocol for digital imaging established by the manufacturer.

      (i) The registrant shall include the protocols established in subparagraph (B) of this paragraph in its operating and safety procedures.

      (ii) The registrant shall document the frequency at which the QA/QC protocol is performed. Documentation shall: