    (D) signature of the authorized medical physicist who performed the full calibration.

(kkk) Full calibration measurements on remote afterloader units.

  (1) A licensee authorized to use a remote afterloader for medical use shall perform full calibration measurements on each unit as follows:

    (A) before the first medical use of the unit;

    (B) before medical use under any of the following conditions:

      (i) following replacement of the sealed source;

      (ii) following reinstallation of the unit in a new location outside the facility; and

      (iii) following any repair of the unit that includes removal of the sealed source or major repair of the components associated with the sealed source exposure assembly;

    (C) at intervals not to exceed three months for high dose-rate, medium dose-rate, and pulsed dose-rate remote afterloader units with sealed sources whose half-life exceeds 75 days; and

    (D) at intervals not to exceed one year for low dose-rate afterloader units.

  (2) Full calibration measurements shall include, as applicable, determination of the following:

    (A) the output within plus or minus five percent;

    (B) sealed source positioning accuracy to within plus or minus 1 millimeter (mm);

    (C) sealed source retraction with backup battery upon power failure;

    (D) length of the sealed source transfer tubes;

    (E) timer accuracy and linearity over the typical range of use;

    (F) length of the applicators; and

    (G) function of the sealed source transfer tubes, applicators, and transfer tube-applicator interfaces.

  (3) A licensee shall use the dosimetry system described in subsection (iii)(1) of this section to measure the output.

  (4) A licensee shall make full calibration measurements required by paragraph (1) of this subsection in accordance with published protocols accepted by nationally recognized bodies.

  (5) In addition to the requirements for full calibrations for low dose-rate remote afterloader units in paragraph (2) of this subsection, a licensee shall perform an autoradiograph of the sealed source(s) to verify inventory and sealed source(s) arrangement at intervals not to exceed three months.

  (6) For low dose-rate remote afterloader units, a licensee may use measurements provided by the sealed source manufacturer that are made in accordance with paragraphs (1) - (5) of this subsection.

  (7) The licensee shall mathematically correct the outputs determined in paragraph (2)(A) of this subsection for physical decay at intervals consistent with one percent physical decay.

  (8) Full calibration measurements required by paragraph (1) of this subsection and physical decay corrections required by paragraph (7) of this subsection shall be performed by an authorized medical physicist.

  (9) The licensee shall retain a record of each calibration in accordance with subsection (xxx) of this section for inspection by the department. The record shall include the following:

    (A) complete date of the calibration including the month, day, and year;

    (B) manufacturer's name, model number and serial number of the remote afterloader unit's sealed source, and the instruments used to calibrate the unit;

    (C) results and an assessment of the full calibrations;

    (D) signature of the authorized medical physicist of this section; and

    (E) results of the autoradiograph required for low dose-rate remote afterloader unit.

(lll) Full calibration measurements on gamma stereotactic radiosurgery units.

  (1) A licensee authorized to use a gamma stereotactic radiosurgery unit for medical use shall perform full calibration measurements on each gamma stereotactic radiosurgery unit as follows:

    (A) before the first medical use of the unit;

    (B) before medical use under the following conditions:

      (i) whenever spot check measurements indicate that the output differs by more than five percent from the output obtained at the last full calibration corrected mathematically for radioactive decay;

      (ii) following replacement of the sealed sources or following reinstallation of the gamma stereotactic radiosurgery unit in a new location; and

      (iii) following any repair of the gamma stereotactic radiosurgery unit that includes removal of the sealed sources or major repair of the components associated with the sealed source exposure assembly; and

    (C) at intervals not to exceed one year, with the exception that relative helmet factors need only be determined before the first medical use of a helmet and following any damage to a helmet.

  (2) Full calibration measurements shall include determination of the following:

    (A) the output within plus or minus three percent;

    (B) relative helmet factors;

    (C) isocenter coincidence;

    (D) timer accuracy and linearity over the range of use;

    (E) "on-off" error;

    (F) trunnion centricity;

    (G) treatment table retraction mechanism, using backup battery power or hydraulic backups with the unit "off";

    (H) helmet microswitches;

    (I) emergency timing circuits; and

    (J) stereotactic frames and localizing devices (trunnions).

  (3) The licensee shall use the dosimetry system described in subsection (iii)(1) of this section to measure the output for one set of exposure conditions. The remaining radiation measurements required in paragraph (2)(A) of this subsection may be made using a dosimetry system that indicates relative dose rates.

  (4) The licensee shall make full calibration measurements required by paragraph (1) of this subsection in accordance with published protocols accepted by nationally recognized bodies.

  (5) The licensee shall mathematically correct the outputs determined in paragraph (2)(A) of this subsection at intervals not to exceed one month for cobalt-60 and at intervals consistent with one percent physical decay for all other radionuclides.

  (6) Full calibration measurements required by paragraph (1) of this subsection and physical decay corrections required by paragraph (5) of this subsection shall be performed by an authorized medical physicist.

  (7) The licensee shall retain a record of each calibration in accordance with subsection (xxx) of this section for inspection by the department. The record shall include the following:

    (A) complete date of the calibration including the month, day and year;

    (B) manufacturer's name, model number, and serial number for the unit and the unit's sealed source and the instruments used to calibrate the unit;

    (C) results and an assessment of the full calibration; and

    (D) signature of the authorized medical physicist who performed the full calibration.

(mmm) Periodic spot checks for teletherapy units.

  (1) A licensee authorized to use teletherapy units for medical use shall perform output spot checks on each teletherapy unit once in each calendar month that include determination of the following:

    (A) timer constancy and linearity over the range of use;

    (B) "on-off" error;

    (C) the coincidence of the radiation field and the field indicated by the light beam localizing device;

    (D) the accuracy of all distance measuring and localization devices used for medical use;

    (E) the output for one typical set of operating conditions measured with the dosimetry system described in subsection (iii)(2) of this section; and

    (F) the difference between the measurement made in subparagraph (E) of this paragraph and the anticipated output, expressed as a percentage of the anticipated output, the value obtained at last full calibration corrected mathematically for physical decay.

  (2) The licensee shall perform measurements required by paragraph (1) of this subsection in accordance with written procedures established by an authorized medical physicist. That authorized medical physicist need not actually perform the spot check measurements. The licensee shall maintain a copy of the written procedures in accordance with subsection (xxx) of this section for inspection by the department.

  (3) The licensee authorized to use a teletherapy unit for medical use shall perform safety spot checks of each teletherapy facility once in each calendar month and after each sealed source installation to assure proper operation of the following: