<<Prev Rule

Texas Administrative Code

Next Rule>>
TITLE 25HEALTH SERVICES
PART 1DEPARTMENT OF STATE HEALTH SERVICES
CHAPTER 289RADIATION CONTROL
SUBCHAPTER EREGISTRATION REGULATIONS
RULE §289.229Radiation Safety Requirements for Accelerators, Therapeutic Radiation Machines, Simulators, and Electronic Brachytherapy Devices

      (i) Each installation shall be provided with primary and/or secondary barriers as are necessary to assure compliance with §289.231(m) and (o) of this title. All protective barriers shall be fixed except for entrance doors or beam interceptors.

      (ii) The control panel shall be located outside the treatment room or in an enclosed booth inside the room.

      (iii) Interlocks shall be provided such that all entrance doors shall be closed, including doors to any interior booths, before treatment can be initiated or continued. If the radiation beam is interrupted by any door opening, it shall not be possible to restore the machine to operation without closing the door and reinitiating irradiation by manual action at the control panel. When any door is opened while the x-ray tube is activated, the exposure at a distance of 1 m from the source shall be reduced to less than 100 mR/hr (1 mGy/hr).

    (D) Surveys, calibrations, and spot checks.

      (i) Surveys shall be performed as follows.

        (I) All new and existing facilities not previously surveyed shall have an initial survey made by a licensed medical physicist with a specialty in therapeutic radiological physics, who shall provide a written report of the survey to the registrant. Additional surveys shall be done after any change in the facility, facility design, or equipment that might cause a significant increase in radiation hazard.

        (II) The registrant shall maintain a copy of the initial survey report and all subsequent survey reports required by subclause (I) of this clause in accordance with subsection (l) of this section for inspection by the agency.

        (III) The survey report shall indicate all instances where the installation is in violation of applicable requirements of this chapter.

      (ii) Calibrations shall be performed as follows.

        (I) The calibration of a therapeutic radiation system shall be performed at intervals not to exceed 1 year and after any change or replacement of components that could cause a change in the radiation output. The calibrations shall be such that the dose at a reference point in a water or plastic phantom can be calculated to within an uncertainty of 5.0%.

        (II) The calibration of the radiation output of the therapeutic radiation system shall be performed by a licensed medical physicist with a specialty in therapeutic radiological physics who is physically present at the facility during such calibration.

        (III) The calibration of the therapeutic radiation system shall include, but not be limited to, the following determinations:

          (-a-) verification that the radiation therapy system is operating in compliance with the design specifications;

          (-b-) HVL for each kV setting and filter combination used;

          (-c-) the exposure rates (air kerma rates) as a function of field size, technique factors, filter, and treatment distance used; and

          (-d-) the degree of congruence between the radiation field and the field indicated by the localizing device, if such device is present, which shall be within 5 mm for any field edge.

        (IV) Calibration of the radiation output of a therapeutic radiation system shall be performed with a calibrated dosimetry system. The dosimetry system calibration shall be traceable to a national standard. The calibration interval shall not exceed 24 months.

        (V) Records of calibration measurements specified in clause (ii) of this subparagraph shall be maintained by the registrant in accordance with subsection (l) of this section for inspection by the agency.

        (VI) A copy of the latest calibrated absorbed dose rate measured on a particular therapeutic radiation system shall be available at a designated area within the therapy facility housing that therapeutic radiation system.

      (iii) Spot checks shall be performed on therapeutic radiation systems capable of operation at greater than 150 kVp. Such measurements shall meet the following requirements.

        (I) The spot check procedures shall be in writing, or documented in an electronic reporting system, and shall have been developed by a licensed medical physicist with a specialty in therapeutic radiological physics.

        (II) If a licensed medical physicist does not perform the spot check measurements, the results of the spot check measurements shall be reviewed by a licensed medical physicist with a specialty in therapeutic radiological physics within 5 treatment days and a record made of the review. If the output varies by more than 5.0% from the expected value, a licensed medical physicist with a specialty in therapeutic radiological physics shall be notified immediately.

        (III) The written spot check procedures shall specify the frequency that tests or measurements are to be performed and that the spot check shall be performed during the calibration specified in clause (ii) of this subparagraph. The acceptable tolerance for each parameter measured when compared to the value for that parameter determined in the calibration specified in clause (ii) of this subparagraph shall be stated.

        (IV) The written spot check procedures shall include special operating instructions that shall be carried out whenever a parameter in subclause (III) of this clause exceeds an acceptable tolerance.

        (V) Whenever a spot check indicates a significant change in the operating characteristics of a system, as specified in the procedures, the system shall be recalibrated, as required in clause (ii) of this subparagraph.

        (VI) Records of written spot checks and any necessary corrective actions shall be maintained by the registrant in accordance with subsection (l) of this section for inspection by the agency. A copy of the most recent spot check shall be available at a designated area within the therapy facility housing that therapeutic radiation system.

        (VII) Spot checks shall be obtained using a system satisfying the requirements of clause (ii)(IV) of this subparagraph.

  (3) Therapeutic radiation machines capable of operating at energies of 1 MeV and above.

    (A) Equipment requirements.

      (i) For operating conditions producing maximum leakage radiation, the absorbed dose in rads (mGy) due to leakage radiation, including x rays, electrons, and neutrons, at any point in a circular plane of 2 m radius centered on and perpendicular to the central axis of the beam at the isocenter or normal treatment distance and outside the maximum useful beam size shall not exceed 0.1% of the maximum absorbed dose in rads (mGy) of the unattenuated useful beam measured at the point of intersection of the central axis of the beam and the plane surface. Measurements excluding those for neutrons shall be averaged over an area up to, but not exceeding, 100 square centimeters (cm2 ) at the positions specified. Measurements of the portion of the leakage radiation dose contributed by neutrons shall be averaged over an area up to, but not exceeding, 200 cm2 . For each system, the registrant shall determine or obtain from the manufacturer the leakage radiation existing at the positions specified for the specified operating conditions. Records on leakage radiation measurements shall be maintained in accordance with subsection (l) of this section for inspection by the agency.

      (ii) Each wedge filter that is removable from the system shall be clearly marked with an identification number. Documentation available at the control panel shall contain a description of the filter. The wedge angle shall appear on the wedge or wedge tray (if permanently mounted to the tray). If the wedge tray is damaged, the wedge transmission factor shall be redetermined. Equipment manufactured after March 1, 1989, shall meet the following requirements.

        (I) Irradiation shall not be possible until a selection of a filter or a positive selection to use "no filter" has been made at the treatment console, either manually or automatically.

        (II) An interlock system shall be provided to prevent irradiation if the filter selected is not in the correct position.

        (III) A display shall be provided at the treatment console showing the beam quality in use.

        (IV) An interlock shall be provided to prevent irradiation if any filter selection operation carried out in the treatment room does not agree with the filter selection operation carried out at the treatment console.

      (iii) The registrant shall determine data sufficient to assure that the following beam quality requirements in tissue equivalent material are met.

        (I) The absorbed dose resulting from x rays in a useful electron beam at a point on the central axis of the beam 10 cm greater than the practical range of the electrons shall not exceed the values stated in the following Table II. Linear interpolation shall be used for values not stated.

Attached Graphic

        (II) Compliance with subclause (I) of this clause shall be determined using:

Cont'd...

Next Page Previous Page

Link to Texas Secretary of State Home Page | link to Texas Register home page | link to Texas Administrative Code home page | link to Open Meetings home page