<<Prev Rule

Texas Administrative Code

Next Rule>>
TITLE 25HEALTH SERVICES
PART 1DEPARTMENT OF STATE HEALTH SERVICES
CHAPTER 289RADIATION CONTROL
SUBCHAPTER DGENERAL
RULE §289.202Standards for Protection Against Radiation from Radioactive Materials

      (i) the radiation level from the source of radiation is reduced below that at which it would be possible for an individual to receive a deep dose equivalent in excess of 0.1 rem (1 mSv) in one hour; and

      (ii) conspicuous visible and audible alarm signals are generated to make potentially affected individuals aware of the hazard and the licensee or at least one other individual, who is familiar with the activity and prepared to render or summon assistance, aware of the failure or removal of the physical barrier.

    (D) When the shield for stored sealed sources is a liquid, the licensee shall provide means to monitor the integrity of the shield and to signal, automatically, loss of adequate shielding.

    (E) Physical radiation barriers that comprise permanent structural components, such as walls, that have no credible probability of failure or removal in ordinary circumstances, need not meet the requirements of subparagraphs (C) and (D) of this paragraph.

    (F) Each area shall be equipped with devices that will automatically generate conspicuous visible and audible alarm signals to alert personnel in the area before the source of radiation can be put into operation and in time for any individual in the area to operate a clearly identified control device, which must be installed in the area and which can prevent the source of radiation from being put into operation.

    (G) Each area shall be controlled by use of such administrative procedures and such devices as are necessary to ensure that the area is cleared of personnel prior to each use of the source of radiation.

    (H) Each area shall be checked by a radiation measurement to ensure that, prior to the first individual's entry into the area after any use of the source of radiation, the radiation level from the source of radiation in the area is below that at which it would be possible for an individual to receive a deep dose equivalent in excess of 0.1 rem (1 mSv) in one hour.

    (I) The entry control devices required in subparagraph (A) of this paragraph shall be tested for proper functioning. See subsection (uu) of this section for recordkeeping requirements.

      (i) Testing shall be conducted prior to initial operation with the source of radiation on any day, unless operations were continued uninterrupted from the previous day.

      (ii) Testing shall be conducted prior to resumption of operation of the source of radiation after any unintentional interruption.

      (iii) The licensee shall submit and adhere to a schedule for periodic tests of the entry control and warning systems.

    (J) The licensee shall not conduct operations, other than those necessary to place the source of radiation in safe condition or to effect repairs on controls, unless control devices are functioning properly.

    (K) Entry and exit portals that are used in transporting materials to and from the irradiation area, and that are not intended for use by individuals, shall be controlled by such devices and administrative procedures as are necessary to physically protect and warn against inadvertent entry by any individual through these portals. Exit portals for irradiated materials shall be equipped to detect and signal the presence of any loose radioactive material that is carried toward such an exit and automatically to prevent loose radioactive material from being carried out of the area.

  (3) Licensees or applicants for licenses for sources of radiation within the purview of paragraph (2) of this subsection that will be used in a variety of positions or in locations, such as open fields or forests, which make it impracticable to comply with certain requirements of paragraph (2) of this subsection, such as those for the automatic control of radiation levels, may apply to the agency for approval of alternative safety measures. Alternative safety measures shall provide personnel protection at least equivalent to those specified in paragraph (2) of this subsection. At least one of the alternative measures shall include an entry-preventing interlock control based on a measurement of the radiation that ensures the absence of high radiation levels before an individual can gain access to the area where such sources of radiation are used.

  (4) The entry control devices required by paragraphs (2) and (3) of this subsection shall be established in such a way that no individual will be prevented from leaving the area.

(v) Use of process or other engineering controls. The licensee shall use, to the extent practicable, process or other engineering controls, such as containment, decontamination, or ventilation, to control the concentrations of radioactive material in air.

(w) Use of other controls.

  (1) When it is not practicable to apply process or other engineering controls to control the concentrations of radioactive material in air to values below those that define an airborne radioactivity area, the licensee shall, consistent with maintaining the total effective dose equivalent ALARA, increase monitoring and limit intakes by one or more of the following means:

    (A) control of access;

    (B) limitation of exposure times;

    (C) use of respiratory protection equipment; or

    (D) other controls.

  (2) If the licensee performs an ALARA analysis to determine whether respirators should be used, the licensee may consider safety factors other than radiological factors. The licensee shall also consider the impact of respirator use on workers' industrial health and safety.

(x) Use of individual respiratory protection equipment.

  (1) If the licensee uses respiratory protection equipment to limit intakes of radioactive material in accordance with subsection (w) of this section, the licensee shall do the following.

    (A) Except as provided in subparagraph (B) of this paragraph, the licensee shall use only respiratory protection equipment that is tested and certified by the National Institute for Occupational Safety and Health (NIOSH).

    (B) If the licensee wishes to use equipment that has not been tested or certified by the NIOSH, or for which there is no schedule for testing or certification, the licensee shall submit an application to the agency for authorized use of that equipment, including a demonstration by testing, or a demonstration on the basis of test information, that the material and performance characteristics of the equipment are capable of providing the proposed degree of protection under anticipated conditions of use.

    (C) The licensee shall implement and maintain a respiratory protection program that includes:

      (i) air sampling sufficient to identify the potential hazard, permit proper equipment selection, and estimate doses;

      (ii) surveys and bioassays, as appropriate, to evaluate actual intakes;

      (iii) testing of respirators for operability (user seal check for face sealing devices and functional check for others) immediately prior to each use;

      (iv) written procedures regarding the following:

        (I) monitoring, including air sampling and bioassays;

        (II) supervision and training of respirator users;

        (III) fit testing;

        (IV) respirator selection;

        (V) breathing air quality;

        (VI) inventory and control;

        (VII) storage, issuance, maintenance, repair, testing, and quality assurance of respiratory protection equipment;

        (VIII) recordkeeping; and

        (IX) limitations on periods of respirator use and relief from respirator use;

      (v) determination by a physician prior to initial fitting of a face sealing respirator and the first field use of non-face sealing respirators, and either every 12 months thereafter or periodically at a frequency determined by a physician, that the individual user is medically fit to use the respiratory protection equipment; and

      (vi) fit testing, with fit factor >10 times the APF for negative pressure devices, and a fit factor > 500 for any positive pressure, continuous flow, and pressure-demand devices, before the first field use of tight fitting, face-sealing respirators and periodically thereafter at a frequency not to exceed 1 year. Fit testing shall be performed with the facepiece operating in the negative pressure mode.

    (D) The licensee shall advise each respirator user that the user may leave the area at any time for relief from respirator use in the event of equipment malfunction, physical or psychological distress, procedural or communication failure, significant deterioration of operating conditions, or any other conditions that might require such relief.

Cont'd...

Next Page Previous Page

Link to Texas Secretary of State Home Page | link to Texas Register home page | link to Texas Administrative Code home page | link to Open Meetings home page