<<Prev Rule

Texas Administrative Code

Next Rule>>
TITLE 25HEALTH SERVICES
PART 1DEPARTMENT OF STATE HEALTH SERVICES
CHAPTER 289RADIATION CONTROL
SUBCHAPTER EREGISTRATION REGULATIONS
RULE §289.227Use of Radiation Machines in the Healing Arts

      (i) The fluoroscopic imaging assembly shall be provided with a primary protective barrier that intercepts the entire cross section of the useful beam at any SID.

      (ii) The x-ray tube used for fluoroscopy shall not produce x-rays unless the barrier is in position to intercept the useful beam and the imaging device is in place and operable.

      (iii) The exposure rate (air kerma rate) due to transmission through the barrier with the attenuation block in the useful beam, combined with radiation through the image intensifier if provided, shall not exceed 3.34 x 10-3 % of the entrance exposure rate (air kerma rate) at a distance of 10 cm from any accessible surface of the fluoroscopic imaging assembly beyond the plane of the image receptor.

    (B) Measuring compliance of barrier transmission.

      (i) The exposure rate (air kerma rate) due to transmission through the primary protective barrier combined with radiation through the image intensifier shall be determined by measurements averaged over an area of 100 cm2 with no linear dimension greater than 20 cm.

      (ii) If the source is below the tabletop, the measurement shall be made with the input surface of the fluoroscopic imaging assembly positioned 30 cm above the tabletop.

      (iii) If the source is above the tabletop and the SID is variable, the measurement shall be made with the end of the beam-limiting device or spacer as close to the tabletop as it can be placed, provided that it shall not be closer than 30 cm.

      (iv) Movable grids and compression devices shall be removed from the useful beam during the measurement.

      (v) The attenuation block shall be positioned in the useful beam 10 cm from the point of measurement of entrance exposure rate (air kerma rate) and between this point and the input surface of the fluoroscopic imaging assembly.

      (vi) The collimator shall be fully open when the measurement is made.

    (C) X-ray field.

      (i) Compliance with clauses (ii) - (vi) of this subparagraph shall be determined with the beam axis perpendicular to the plane of the image receptor.

      (ii) Fluoroscopic systems with a fixed SID and the capability of a visible area of no greater than 300 cm2 shall be provided with either stepless adjustment of the x-ray field or a means to further limit the x-ray field at the image receptor to 125 cm2 or less. If the equipment is provided with stepless adjustment, the minimum x-ray field size at the maximum SID shall be less than or equal to 5 cm by 5 cm at the image receptor.

      (iii) Fluoroscopic systems with a variable SID or a fixed SID with the capability of a visible area of greater than 300 cm2 shall be provided with stepless adjustment of the field size. The minimum x-ray field size at the maximum SID shall be less than or equal to 5 cm by 5 cm at the image receptor.

      (iv) Neither the length nor the width of the x-ray field in the plane of the image receptor shall exceed that of the visible area of the image receptor by more than 3.0% of the SID. The sum of the excess length and the excess width shall be no greater than 4.0% of the SID.

      (v) For rectangular x-ray fields used with circular image receptors, the error in alignment shall be determined along the length and width dimensions of the x-ray field that pass through the center of the visible area of the image receptor.

      (vi) For fluoroscopic systems with only a manual mode of collimation, the x-ray field produced shall be limited to the area of the spot-film cassette at 16 inches above tabletop. Additionally, during fluoroscopy, the beam shall be restricted to the area of the input phosphor.

      (vii) Spot-film devices shall meet the following additional requirements.

        (I) Means shall be provided between the source and the patient for adjustment of the x-ray field size in the plane of the film to the size of that portion of the film that has been selected on the spot-film selector.

          (-a-) Such adjustment shall be automatically accomplished except when the x-ray field size in the plane of the film is smaller than that of the selected portion of the film.

          (-b-) The total misalignment of the edges of the x-ray field with the respective edges of the selected portion of the image receptor along the length or width dimensions of the x-ray field in the plane of the image receptor shall not exceed 3.0% of the SID when adjusted for full coverage of the selected portion of the image receptor.

          (-c-) The sum, without regard to sign of the misalignment along any two orthogonal dimensions, shall not exceed 4.0% of the SID.

        (II) The center of the x-ray field in the plane of the film shall be aligned with the center of the selected portion of the film to within plus or minus 2.0% of the SID.

  (2) Activation of the fluoroscopic tube. X-ray production in the fluoroscopic mode shall be controlled by a device that requires continuous pressure by the fluoroscopist for the entire time of the exposure (continuous pressure type switch). When recording serial fluoroscopic images, the fluoroscopist shall be able to terminate the x-ray exposures at any time, but means may be provided to permit completion of any single exposure of the series in process.

  (3) Entrance exposure rate (air kerma rate) allowable limits for fluoroscopic systems.

    (A) The following requirements apply to fluoroscopic systems manufactured prior to May 19, 1995.

      (i) Fluoroscopic systems with AERC. Fluoroscopic systems that are provided with AERC shall not be operable at any combination of tube potential and current that will result in an exposure rate (air kerma rate) in excess of 2.58 x 10-3 coulomb per kilogram per minute (C/kg/min) (10 roentgens per minute (10 R/min) or (100 mGy/min) at the point where the center of the useful beam enters the patient, except:

        (I) during recording of fluoroscopic images, excluding last image hold; or

        (II) when an optional high-level control is provided. When so provided, the fluoroscopic system shall not be operable at any combination of tube potential and current that will result in an exposure rate (air kerma rate) in excess of 1.29 x 10-3 C/kg/min (5 R/min or 50 mGy/min) at the point where the center of the useful beam enters the patient, unless the high-level control is activated. Special means of activation of high-level controls shall be required. The high-level control shall be operable only when continuous manual activation is provided by the operator. A continuous signal audible to the fluoroscopist shall indicate that the high-level control is being employed.

      (ii) Fluoroscopic systems without AERC (manual mode). Fluoroscopic systems that are not provided with AERC shall not be operable at any combination of tube potential and current that will result in an exposure rate (air kerma rate) in excess of 1.29 x 10-3 C/kg/min (5 R/min or 50 mGy/min) at the point where the center of the useful beam enters the patient, except:

        (I) during recording of fluoroscopic images, excluding last image hold; or

        (II) when an optional high-level control is activated. Special means of activation of high-level controls shall be required. The high-level control shall be operable only when continuous manual activation is provided by the operator. A continuous signal audible to the fluoroscopist shall indicate that the high-level control is being employed.

      (iii) Fluoroscopic systems with both an AERC mode and a manual mode. Fluoroscopic systems that are provided with both an AERC mode and a manual mode shall not be operable at any combination of tube potential and current that will result in an exposure rate (air kerma rate) in excess of 2.58 x 10-3 C/kg/min (10 R/min or 100 mGy/min) in either mode at the point where the center of the useful beam enters the patient except:

        (I) during recording of fluoroscopic images, excluding last image hold; or

        (II) when the mode or modes have an optional high-level control, in which case that mode or modes shall not be operable at any combination of tube potential and current that will result in an exposure rate (air kerma rate) in excess of 1.29 x 10-3 C/kg/min (5 R/min or 50 mGy/min) at the point where the center of the useful beam enters the patient, unless the high level control is activated. Special means of activation of high-level control shall be required. The high level control shall be operable only when continuous manual activation is provided by the operator. A continuous signal audible to the fluoroscopist shall indicate that the high-level is being employed.

      (iv) Measuring compliance. Compliance of entrance exposure rate (air-kerma rate) for fluoroscopic systems shall be determined as follows.

        (I) If the source is below the x-ray table, the exposure rate (air kerma rate) shall be measured at 1 cm above the tabletop or cradle.

Cont'd...

Next Page Previous Page

Link to Texas Secretary of State Home Page | link to Texas Register home page | link to Texas Administrative Code home page | link to Open Meetings home page