TITLE 25HEALTH SERVICES
PART 1DEPARTMENT OF STATE HEALTH SERVICES
CHAPTER 289RADIATION CONTROL
SUBCHAPTER FLICENSE REGULATIONS
RULE §289.252Licensing of Radioactive Material

    (E) details of quality control procedures to assure that production sources and devices meet the standards of the design and prototype tests;

    (F) procedures and standards for calibrating sealed sources and devices;

    (G) instructions for handling and storing the sealed source or device from the radiation safety standpoint. These instructions are to be included on a durable label attached to the sealed source or device or attached to a permanent storage container for the sealed source or device, provided that instructions that are too lengthy for the label may be summarized on the label and printed in detail on a brochure that is referenced on the label; and

    (H) a legend and methods for labeling sources and devices as to their radioactive content;

  (2) documentation that the label affixed to the sealed source or device, or to the permanent storage container for the sealed source or device, contains information on the radionuclide, quantity, and date of assay, and a statement that the name of the sealed source or device is licensed by the department for commercial distribution to persons licensed for use of sealed sources in the healing arts or by equivalent licenses of the NRC or any agreement state;

  (3) documentation that in the event the applicant desires that the sealed source or device be required to be tested for radioactive material leakage at intervals longer than 6 months, the applicant shall include in the application sufficient information to demonstrate that the longer interval is justified by performance characteristics of the sealed source or device or similar sources or devices and by design features that have a significant bearing on the probability or consequences of radioactive material leakage from the sealed source;

  (4) documentation that in determining the acceptable interval for testing radioactive material leakage, information will be considered that includes the following:

    (A) primary containment (sealed source capsule);

    (B) protection of primary containment;

    (C) method of sealing containment;

    (D) containment construction materials;

    (E) form of contained radioactive material;

    (F) maximum temperature withstood during prototype tests;

    (G) maximum pressure withstood during prototype tests;

    (H) maximum quantity of contained radioactive material;

    (I) radiotoxicity of contained radioactive material; and

    (J) operating experience with identical sealed sources or devices or similarly designed and constructed sealed sources or devices; and

  (5) the source or device has been registered in the Sealed Source and Device Registry.

(p) Specific licenses for the manufacture and commercial distribution of radioactive material for certain in vitro clinical or laboratory testing in accordance with the general license. In addition to the requirements in subsection (e) of this section, a specific license to manufacture or commercially distribute radioactive material for use in accordance with the general license in §289.251(f)(4)(G) of this title will be issued if the department approves the following information submitted by the applicant:

  (1) documentation that the radioactive material will be prepared for distribution in prepackaged units of:

    (A) iodine-125 in units not exceeding 10 µCi (0.37 megabecquerel (MBq)) each;

    (B) iodine-131 in units not exceeding 10 µCi (0.37 MBq) each;

    (C) carbon-14 in units not exceeding 10 µCi (0.37 MBq) each;

    (D) hydrogen-3 (tritium) in units not exceeding 50 µCi (1.85 MBq) each;

    (E) iron-59 in units not exceeding 20 µCi (0.74 MBq) each;

    (F) cobalt-57 in units not exceeding 10 µCi (0.37 MBq) each;

    (G) selenium-75 in units not exceeding 10 µCi (0.37 MBq) each; or

    (H) mock iodine-125 in units not exceeding 0.05 µCi (1.85 kBq) of iodine-129 and 0.005 µCi (0.185 kBq) of americium-241 each;

  (2) evidence that each prepackaged unit will bear a durable, clearly visible label:

    (A) identifying the radioactive contents as to chemical form and radionuclide, and indicating that the amount of radioactivity does not exceed 10 µCi (0.37 MBq) of iodine-125, iodine-131, carbon-14, cobalt-57, or selenium-75; 50 µCi (1.85 MBq) of hydrogen-3 (tritium); 20 µCi (0.74 MBq) of iron-59; or mock iodine-125 in units not exceeding 0.05 µCi (1.85 kBq) of iodine-129 and 0.005 µCi (0.185 kBq) of americium-241; and

    (B) displaying the radiation caution symbol in accordance with §289.202(z) of this title and the words, "CAUTION, RADIOACTIVE MATERIAL," and "Not for Internal or External Use in Humans or Animals";

  (3) that one of the following statements, as appropriate, or a substantially similar statement appears on a label affixed to each prepackaged unit or appears in a leaflet or brochure that accompanies the package:

    (A) option 1:

Attached Graphic

    (B) option 2:

Attached Graphic

  (4) that the label affixed to the unit, or the leaflet or brochure that accompanies the package, contains adequate information as to the precautions to be observed in handling and storing the radioactive material. In the case of a mock iodine-125 reference or calibration source, the information accompanying the source shall also contain directions to the licensee regarding the waste disposal requirements of §289.202(ff) of this title.

(q) Specific licenses for the manufacture and commercial distribution of ice detection devices. In addition to the requirements of subsection (e) of this section, a specific license to manufacture and commercially distribute ice detection devices to persons generally licensed in accordance with §289.251(f)(4)(E) of this title will be issued if the department approves the information submitted by the applicant. This information shall satisfy the requirements of Title 10, CFR, §§32.61 and 32.62.

(r) Specific licenses for the manufacture, preparation, or transfer for commercial distribution of radioactive drugs containing radioactive materials for medical use under §289.256 of this title.

  (1) In addition to the requirements in subsection (e) of this section, a specific license to manufacture, prepare, or transfer for commercial distribution, radioactive drugs containing radioactive material for use by persons authorized in accordance with §289.256 of this title will be issued if the department approves the following information submitted by the applicant:

    (A) evidence that the applicant is at least one of the following:

      (i) registered with the United States Food and Drug Administration (FDA) as the owner or operator of a drug establishment that engages in the manufacture, preparation, propagation, compounding, or processing of a drug in accordance with Title 21, CFR, §207.17;

      (ii) registered or licensed with a state agency as a drug manufacturer;

      (iii) licensed as a pharmacy by the Texas State Board of Pharmacy;

      (iv) operating as a nuclear pharmacy within a federal medical institution; or

      (v) a positron emission tomography (PET) drug production facility registered with a state agency;

    (B) radionuclide data relating to the following:

      (i) chemical and physical form;

      (ii) maximum activity per vial, syringe, generator, or other container of the radioactive drug; and

      (iii) shielding provided by the packaging to show it is appropriate for the safe handling and storage of the radioactive drugs by medical use licensees;

    (C) labeling requirements including the following:

      (i) that each transport radiation shield, whether it is constructed of lead, glass, plastic, or other material, of a radioactive drug to be transferred for commercial distribution shall include the following:

        (I) the radiation symbol and the words "CAUTION, RADIOACTIVE MATERIAL" or "DANGER, RADIOACTIVE MATERIAL;"

        (II) the name of the radioactive drug or its abbreviation; and

        (III) the quantity of radioactivity at a specified date and time (the time may be omitted for radioactive drugs with a half-life greater than 100 days); and

      (ii) that each syringe, vial, or other container used to hold a radioactive drug to be transferred for commercial distribution shall include the following:

        (I) radiation symbol and the words, "CAUTION, RADIOACTIVE MATERIAL" or "DANGER, RADIOACTIVE MATERIAL;" and

Cont'd...