<<Prev Rule

Texas Administrative Code

Next Rule>>
TITLE 25HEALTH SERVICES
PART 1DEPARTMENT OF STATE HEALTH SERVICES
CHAPTER 289RADIATION CONTROL
SUBCHAPTER EREGISTRATION REGULATIONS
RULE §289.227Use of Radiation Machines in the Healing Arts

        (II) If the source is above the x-ray table, the exposure rate (air kerma rate) shall be measured at 30 cm above the tabletop with the end of the beam-limiting device or spacer positioned as closely as possible to the point of measurement.

        (III) In a C-arm fluoroscopic system, the exposure rate (air kerma rate) shall be measured at 30 cm from the input surface of the fluoroscopic imaging assembly provided that the end of the beam-limiting device or spacer is no closer than 30 cm from the input surface of the fluoroscopic imaging assembly. The applicable limit shall not be exceeded at any available SID.

        (IV) In a lateral (horizontal) fluoroscopic system, the exposure rate (air kerma rate) shall be measured at a point 15 cm from the centerline of the x-ray table and in the direction of the x-ray source with the end of the beam-limiting device or spacer positioned as closely as possible to the point of measurement. If the table top is movable, it shall be positioned as closely as possible to the lateral x-ray source, with the end of the beam-limiting device or spacer no closer than 15 cm to the centerline of the x-ray table.

    (B) The following requirements apply to fluoroscopic systems manufactured on or after May 19, 1995.

      (i) Fluoroscopic systems operable at any combination of tube potential and current that will result in an exposure rate (air kerma rate) greater than 1.29 x 10-3 C/kg/min (5 R/min or 50 mGy/min) at the point where the center of the useful beam enters the patient shall be equipped with AERC. Provision for manual selection of technique factors may be provided.

      (ii) Fluoroscopic systems shall not be operable at any combination of tube potential and current that will result in an exposure rate (air kerma rate) in excess of 2.58 x 10-3 C/kg/min (10 R/min or 100 mGy/min) at the point where the center of the useful beam enters the patient except:

        (I) for fluoroscopic systems manufactured prior to June 10, 2006, during the recording of images from an x-ray image-intensifier tube using photographic film or a video camera when the x-ray source is operated in a pulsed mode;

        (II) for fluoroscopic systems manufactured on and after June 10, 2006, during the recording of images from the x-ray image-intensifier to provide the user with a recorded image(s) after termination of the exposure. Such recording does not include images resulting from a last-image-hold feature that are not recorded; and

        (III) when the high-level control is activated, the fluoroscopic system shall not be operable at any combination of tube potential and current that will result in an exposure rate (air kerma rate) in excess of 5.16 x 10-3 C/kg/min (20 R/min or 200 mGy/min) at the point where the center of the useful beam enters the patient. Special means of activation of high-level controls shall be required. The high-level control shall only be operable when continuous manual activation is provided by the operator. A continuous signal audible to the fluoroscopist shall indicate that the high-level control is being employed.

      (iii) Measuring compliance of entrance exposure rate (air-kerma rate) for fluoroscopic systems shall be determined as follows.

        (I) If the source is below the x-ray table, the exposure rate (air kerma rate) shall be measured at 1 cm above the tabletop or cradle.

        (II) If the source is above the x-ray table, the exposure rate (air kerma rate) shall be measured at 30 cm above the tabletop with the end of the beam-limiting device or spacer positioned as closely as possible to the point of measurement.

        (III) In a C-arm fluoroscopic system, the exposure rate (air kerma rate) shall be measured at 30 cm from the input surface of the fluoroscopic imaging assembly provided that the end of the beam-limiting device or spacer is no closer than 30 cm from the input surface of the fluoroscopic imaging assembly. The applicable limit shall not be exceeded at any available SID.

        (IV) In a lateral (horizontal) fluoroscopic system, the exposure rate (air kerma rate) shall be measured at a point 15 cm from the centerline of the x-ray table and in the direction of the x-ray source with the end of the beam-limiting device or spacer positioned as closely as possible to the point of measurement. If the table top is movable, it shall be positioned as closely as possible to the lateral x-ray source, with the end of the beam-limiting device or spacer no closer than 15 cm to the centerline of the x-ray table.

    (C) For hand-held fluoroscopic systems, the exposure rate (air kerma rate) shall be measured at the point closest to the source.

    (D) Periodic measurement of entrance exposure rate (air kerma rate) for fluoroscopic systems shall be performed as follows by a licensed medical physicist.

      (i) Such measurements shall be made within 30 days of installation and within 30 days after any maintenance of the system that might affect the exposure rate. Thereafter, the measurements shall be made annually or at intervals not to exceed 14 months from the date of the prior measurements.

      (ii) Results of these measurements shall be posted where any fluoroscopist may have ready access to such results while using the fluoroscopic system and maintained in accordance with subsection (s)(1) of this section for inspection by the agency. The measurement results shall be stated in R/min or mGy/min and include the technique factors used in determining such results. The name of the person performing the measurements and the date the measurements were performed shall be included in the results.

      (iii) Conditions of periodic measurement of entrance exposure rate (air kerma rate) are as follows.

        (I) The measurement shall be made in accordance with subparagraph (A)(iv) or (B)(iii) of this paragraph, as applicable.

        (II) Fluoroscopic systems that do not incorporate an AERC shall utilize a milliamperage and kVp typical of the clinical use of the fluoroscopic system. Materials should be placed in the useful beam between the detection and imaging systems when conducting these periodic measurements to protect the imaging system.

        (III) Fluoroscopic systems that do incorporate an AERC shall have sufficient material placed in the useful beam to produce a milliamperage and kVp typical of the clinical use of the fluoroscopic system.

  (4) Measurements of the output rate for fluoroscopic systems. Measurements of the output rate of the fluoroscopic system shall be performed with a calibrated dosimetry system in accordance with subsection (i)(14) of this section.

  (5) Indication of potential and current. During fluoroscopy and cinefluorography, the kV and the mA shall be continuously indicated at the control panel and/or the fluoroscopist's position.

  (6) Source-to-skin distance (SSD).

    (A) Means shall be provided to limit the SSD to the following:

      (i) not less than 38 cm on stationary fluoroscopic systems; and

      (ii) not less than 30 cm on portable fluoroscopic systems.

    (B) For image-intensified fluoroscopic systems intended for specific surgical application that would be prohibited at the SSDs specified in subparagraph (A) of this paragraph, provisions may be made for operation at shorter SSDs, but in no case less than 20 cm.

    (C) For stationary or portable C-arm fluoroscopic systems manufactured on or after June 10, 2006, having a maximum source-to-image receptor distance of less than 45 cm, the following requirements shall be met.

      (i) Means shall be provided to limit the SSD to not less than 19 cm.

      (ii) Such systems will be labeled and used for extremity use only.

      (iii) For those systems intended for specific surgical applications that would be prohibited at the SSD specified in clause (i) of this subparagraph, provisions may be made for operation at a shorter SSD, but in no case less than 10 cm.

    (D) The registrant's written operating and safety procedures shall provide precautionary measures to be adhered to during the use of the shorter SSD, in accordance with manufacturer's precautions, if provided.

    (E) The SSD shall not be less than the FDA approved variance for a specific manufacturer of a hand-held fluoroscope.

  (7) Fluoroscopic system timer and display.

    (A) Fluoroscopic systems manufactured prior to June 10, 2006 shall comply with the following requirements.

      (i) Means shall be provided to preset the cumulative on-time of the fluoroscopic x-ray tube. The maximum cumulative time of the timing device shall not exceed 5 minutes without resetting.

Cont'd...

Next Page Previous Page

Link to Texas Secretary of State Home Page | link to Texas Register home page | link to Texas Administrative Code home page | link to Open Meetings home page