<<Prev Rule

Texas Administrative Code

Next Rule>>
TITLE 25HEALTH SERVICES
PART 1DEPARTMENT OF STATE HEALTH SERVICES
CHAPTER 289RADIATION CONTROL
SUBCHAPTER EREGISTRATION REGULATIONS
RULE §289.229Radiation Safety Requirements for Accelerators, Therapeutic Radiation Machines, Simulators, and Electronic Brachytherapy Devices

      (iv) Windows, mirrors, closed-circuit television, or an equivalent system shall be provided to permit continuous observation of the patient following positioning and during irradiation and shall be so located that the operator may observe the patient from the console.

        (I) Should the viewing system described in clause (iv) of this subparagraph fail or be inoperative, treatment shall not be performed with the unit until the system is restored.

        (II) In a facility that has a primary viewing system by electronic means and an alternate viewing system, should both viewing systems described in clause (iv) of this subparagraph fail or be inoperative, treatment shall not be performed with the unit until one of the systems is restored.

      (v) Provision shall be made for continuous two-way aural communication between the patient and the operator at the console independent of the accelerator. However, where excessive noise levels or treatment requirements make aural communication impractical, other methods of communication shall be used. When this is the case, a description of the alternate method shall be submitted to and approved by the agency.

      (vi) Treatment room entrances shall be provided with a warning light in a readily observable position near the outside of all access doors to indicate when the useful beam is "on."

      (vii) Interlocks shall be provided such that all entrance doors shall be closed before treatment can be initiated or continued. If the radiation beam is interrupted by any door opening, it shall not be possible to restore the machine to operation without closing the door and reinitiating irradiation by manual action at the console.

    (C) Surveys, calibrations, spot checks, and operational requirements.

      (i) Surveys shall be performed as follows.

        (I) All new and existing facilities not previously surveyed shall have an initial survey made by a licensed medical physicist with a specialty in therapeutic radiological physics, who shall provide a written report of the survey to the registrant. The physicist who performs the survey shall be a person who did not consult in the design of the therapeutic radiation machine installation and is not employed by or within any corporation or partnership with the person who consulted in the design of the installation. In addition, such surveys shall be done after any change in the facility or equipment that might cause a significant increase in radiation hazard.

        (II) The survey report shall include, but not be limited to the following:

          (-a-) a diagram of the facility that details building structures and the position of the console, therapeutic radiation machine, and associated equipment;

          (-b-) a description of the therapeutic radiation system, including the manufacturer, model and serial number, beam type, and beam energy;

          (-c-) a description of the instrumentation used to determine radiation measurements, including the date and source of the most recent calibration for each instrument used;

          (-d-) conditions under which radiation measurements were taken; and

          (-e-) survey data including:

            (-1-) projected annual TEDE in areas adjacent to the therapy room; and

            (-2-) a description of workload, use, and occupancy factors employed in determining the projected annual TEDE.

        (III) The registrant shall maintain a copy of the survey report and a copy of the survey report shall be provided to the agency within 30 days of completion of the survey. Records of the survey report shall be maintained in accordance with subsection (l) of this section for inspection by the agency.

        (IV) The survey report shall include documentation of all instances where the installation is in violation of applicable regulations. Any deficiencies detected during the survey shall be corrected prior to using the machine.

      (ii) Calibrations of therapeutic systems shall be performed as follows.

        (I) The calibration of systems subject to this subsection shall be performed in accordance with an established calibration protocol before the system is first used for irradiation of a patient and thereafter at time intervals that do not exceed 12 months and after any change that might significantly alter the calibration, spatial distribution, or other characteristics of the therapy beam. The calibration procedures shall be in writing, or documented in an electronic reporting system, and shall have been developed by a licensed medical physicist with a specialty in therapeutic radiological physics. The calibration protocol entitled, "Protocol for Clinical Reference Dosimetry of High-Energy Photon and Electron Beams," Task Group 51, Radiation Therapy Committee, American Association of Physicists in Medicine, Medical Physics 26(9): 1847 - 1870, September 1999, would be accepted as an established protocol. At a minimum, the calibration protocol shall include items in subclauses (III) - (V) of this clause below.

        (II) The calibration shall be performed by a licensed medical physicist with a specialty in therapeutic radiological physics who is physically present at the facility during the calibration.

        (III) Calibration radiation measurements required by subclause (I) of this clause shall be performed using a dosimetry system:

          (-a-) having a calibration factor for cobalt-60 gamma rays traceable to a national standard;

          (-b-) that is traceable to a national standard and at an interval not to exceed 24 months;

          (-c-) that has been calibrated in such a fashion that an uncertainty can be stated for the radiation quantities monitored by the system; and

          (-d-) that has had constancy checks performed on the system as specified by the licensed medical physicist with a specialty in therapeutic radiological physics.

        (IV) Calibrations shall be in sufficient detail that the dose at a reference point in a tissue equivalent phantom can be calculated to within an uncertainty of 5.0%.

        (V) The calibration of the therapy unit shall include, but not be limited to, the following determinations.

          (-a-) Verification that the equipment is operating in compliance with the design specifications concerning the light field, patient positioning lasers, and back-pointer lights with the isocenter when applicable, variation in the axis of rotation for the table, gantry, and collimator system, and beam flatness and symmetry at the specified depth.

          (-b-) The absorbed dose rate at various depths in a tissue equivalent phantom for the range of field sizes used, for each effective energy, that will verify the accuracy of the dosimetry of all therapy procedures utilized with that therapy beam.

          (-c-) The uniformity of the radiation field to include symmetry, flatness, and dependence on gantry angle.

          (-d-) Verification that existing isodose charts applicable to the specific machine continue to be valid or are updated to existing machine conditions.

          (-e-) Verification of transmission factors for all accessories such as wedges, block trays, and/or universal and custom made beam modifying devices.

        (VI) Calibration of therapeutic systems containing asymmetric jaws, multileaf collimation, or dynamic/virtual wedges shall be performed with an established protocol. The procedures shall be developed by a licensed medical physicist with a specialty in therapeutic radiological physics and shall be in writing or documented in an electronic reporting system. Current recommendations by a national professional association as the American Association of Physicists in Medicine, Task Group 142 report: "Quality Assurance of Medical Accelerators" published August 17, 2009, would be considered an established protocol.

        (VII) Records of calibration measurements specified in subclause (I) of this clause and dosimetry system calibrations specified in subclause (III) of this clause shall be maintained by the registrant in accordance with subsection (l) of this section for inspection by the agency.

        (VIII) A copy of the latest calibrated absorbed dose rate measured in accordance with subclause (I) of this clause shall be available at a designated area within the facility housing that radiation therapy system.

      (iii) Spot checks shall be performed on systems subject to this paragraph during calibrations and thereafter at weekly intervals with the period between spot checks not to exceed 5 treatment days. Such radiation output measurements shall meet the following requirements.

        (I) The spot check procedures shall be performed in accordance with established protocol, shall be in writing, or documented in an electronic reporting system, and shall have been developed by a licensed medical physicist with a specialty in therapeutic radiological physics. Either the spot check protocol entitled, "Comprehensive QA for Radiation Oncology," Task Group 40, Radiation Therapy Committee, American Association of Physicists in Medicine, Medical Physics 21(4): 581-618, Cont'd...

Next Page Previous Page

Link to Texas Secretary of State Home Page | link to Texas Register home page | link to Texas Administrative Code home page | link to Open Meetings home page