<<Prev Rule

Texas Administrative Code

Next Rule>>
TITLE 25HEALTH SERVICES
PART 1DEPARTMENT OF STATE HEALTH SERVICES
CHAPTER 289RADIATION CONTROL
SUBCHAPTER EREGISTRATION REGULATIONS
RULE §289.229Radiation Safety Requirements for Accelerators, Therapeutic Radiation Machines, Simulators, and Electronic Brachytherapy Devices

April, 1994, or Task Group 142 report: Quality Assurance of Medical Accelerators, published by American Association of Physicists in Medicine on August 17, 2009, are accepted as an established protocol. At a minimum, the spot check protocol shall include items in subclauses (III) - (VI) of this clause.

        (II) If a licensed medical physicist does not perform the spot check measurements, the results of the spot check measurements shall be reviewed by a licensed medical physicist at a frequency not to exceed 5 treatment days and a record kept of the review. If the output varies by more than 3.0% from the expected value, a licensed medical physicist shall be notified immediately.

        (III) The written spot check procedures shall specify the frequency at which tests or measurements are to be performed and the acceptable tolerance for each parameter measured in the spot check when compared to the value for that parameter determined in the calibration.

        (IV) Where a system has built-in devices that provide a measurement of any parameter during irradiation, such measurement shall not be utilized as a spot check measurement.

        (V) A parameter exceeding a tolerance set by a licensed medical physicist shall be corrected before the system is used for patient irradiation.

        (VI) Whenever a spot check indicates a significant change in the operating characteristics of a system, as specified in a licensed medical physicist's written procedures, the system shall be recalibrated, as required in this clause of this subparagraph.

        (VII) Records of spot check measurements and any necessary corrective actions shall be maintained by the registrant in accordance with subsection (l) of this section for inspection by the agency.

        (VIII) Spot checks shall be obtained using a system satisfying the requirements of clause (ii)(III) of this subparagraph.

      (iv) Facilities with therapeutic radiation machines with energies of 1 MeV and above shall procure the services of a licensed medical physicist with a specialty in therapeutic radiological physics.

        (I) The physicist shall be responsible for:

          (-a-) calibration of radiation machines;

          (-b-) supervision and review of beam and clinical dosimetry;

          (-c-) measurement, analysis, and tabulation of beam data;

          (-d-) establishment of quality assurance procedures and performance of spot check review; and

          (-e-) review of absorbed doses delivered to patients.

        (II) The licensed medical physicist described in subclause (I) of this clause shall also be available and responsive to immediate problems or emergencies.

  (4) Radiation therapy simulators.

    (A) General requirements. In addition to the general requirements in paragraph (1)(B), (C), (F), and (H) of this subsection, radiation therapy simulators shall comply with the following:

      (i) Technique chart. A technique chart relevant to the particular radiation machine shall be provided or electronically displayed in the vicinity of the console and used by all operators.

      (ii) Operating and safety procedures. Each registrant shall have and implement written operating and safety procedures in accordance with paragraph (1)(G) of this subsection.

      (iii) Protective devices. When utilized, protective devices shall meet the following requirements.

        (I) Protective devices shall be made of no less than 0.25 mm lead equivalent material.

        (II) Protective devices, including aprons, gloves, and shields shall be checked annually for defects, such as holes, cracks, and tears. These checks may be performed by the registrant by visual, tactile, or x-ray imaging. If a defect is found, equipment shall be replaced or removed from service until repaired. A record of this test shall be made and maintained by the registrant in accordance with subsection (l) of this section for inspection by the agency.

      (iv) Viewing system. Windows, mirrors, closed circuit television, or an equivalent system shall be provided to permit the operator to continuously observe the patient during irradiation. The operator shall be able to maintain verbal, visual, and aural contact with the patient.

      (v) Operator position. The operator's position during the exposure shall be such that the operator's exposure is as low as reasonably achievable (ALARA) and the operator is a minimum of 6 feet from the source of radiation or protected by an apron, gloves, or other shielding having a minimum of 0.25 mm lead equivalent material.

      (vi) Holding of tube. In no case shall an individual hold the tube or tube housing assembly supports during any radiographic exposure.

      (vii) No individuals other than the patient and the operator(s) shall be in the treatment room during operation of the simulator.

      (viii) Film processing.

        (I) Films shall be developed in accordance with the time-temperature relationships recommended by the film manufacturer. The specified developer temperature for automatic processing and the time-temperature chart for manual processing shall be posted in the darkroom. If the registrant determines an alternate time-temperature relationship is more appropriate for a specific facility, that time-temperature relationship shall be documented and posted.

        (II) Chemicals shall be replaced according to the chemical manufacturer's or supplier's recommendations or at an interval not to exceed 3 months.

        (III) Darkroom light leak tests shall be performed and any light leaks corrected at intervals not to exceed 6 months.

        (IV) Lighting in the film processing/loading area shall be maintained with the filter, bulb wattage, and distances recommended by the film manufacturer for that film emulsion or with products that provide an equivalent level of protection against fogging.

        (V) Corrections or repairs of the light leaks or other deficiencies in subclauses (II), (III), and (IV) of this clause shall be initiated within 72 hours of discovery and completed no longer than 15 days from detection of the deficiency unless a longer time is authorized by the agency. Records of the correction or repairs shall include the date and initials of the individual performing these functions and shall be maintained in accordance with subsection (l) of this section for inspection by the agency.

        (VI) Documentation of the items in subclauses (II), (III), and (V) of this clause shall be maintained at the site where performed and shall include the date and initials of the individual completing these items. These records shall be kept in accordance with subsection (l) of this section for inspection by the agency.

      (ix) Alternative processing systems. Users of daylight processing systems, laser processors, self-processing film units, or other alternative processing systems shall follow manufacturer's recommendations for image processing. Documentation that the registrant is following manufacturer's recommendations shall include the date and initials of the individual completing the document and shall be maintained at the site where performed in accordance with subsection (l) of this section for inspection by the agency.

      (x) Digital imaging acquisition systems. Users of digital imaging acquisition systems shall follow quality assurance/quality control protocol for image processing established by the manufacturer or, if no manufacturer's protocol is available, by the registrant. The registrant shall include the protocol, whether established by the registrant or the manufacturer in its operating and safety procedures. The registrant shall document the frequency at which the quality assurance/quality control protocol is performed. Documentation shall include the date and initials of the individual completing the document and shall be maintained at the site where performed in accordance with subsection (l) of this section for inspection by the agency.

    (B) Additional requirements for radiation therapy simulators used in the general radiographic mode of operation.

      (i) Beam quality (HVL). The half-value layer of the useful beam for a given x-ray tube potential shall not be less than the values shown in the following Table IV. If it is necessary to determine such half-value layer at an x-ray tube potential that is not listed in Table IV, linear interpolation may be made.

Attached Graphic

      (ii) Technique and exposure indicators.

        (I) The technique factors to be used during an exposure shall be indicated before the exposure begins except when automatic exposure controls are used, in which case the technique factors that are set prior to the exposure shall be indicated.

Cont'd...

Next Page Previous Page

Link to Texas Secretary of State Home Page | link to Texas Register home page | link to Texas Administrative Code home page | link to Open Meetings home page