RULE §289.227Use of Radiation Machines in the Healing Arts

        (I) include a minimum of 1 hour fluoroscopic machine training demonstrating application of clause (i)(I), and (IV) - (VII) of this subparagraph; and

        (II) be provided by a radiologist, licensed medical physicist, or a physician that meets the requirements of this subparagraph.

      (iii) The registrant shall ensure that radiation safety awareness training be completed within 2 years from the effective date of this rule for each physician, "as well as individuals to whom a physician has delegated authority pursuant to the Occupations Code, Chapter 601, and the applicable rules of the Texas Medical Board," performing FGI procedures.

      (iv) Two years after the effective date of this rule, the registrant shall ensure that prior to performing FGI procedures each physician, "as well as individuals to whom a physician has delegated authority pursuant to the Occupations Code, Chapter 601, and the applicable rules of the Texas Medical Board," shall complete the radiation safety awareness training.

      (v) The 8 hours of Category 1 CMEU in radiation safety awareness training required by this subsection may be obtained by web-based online training.

      (vi) The registrant shall maintain radiation safety awareness training records for the 8 hours Category 1 CMEU and the 1 hour fluoroscopic machine training required in this subparagraph in accordance with subsection (s)(1) of this section for inspection by the agency.

(n) CT systems.

  (1) CT system requirements shall include the following.

    (A) Warning label. The warning label will meet the requirements of §289.231(z) of this title.

    (B) Visual indication. The CT control panel shall provide visual indication of the production of x-rays.

    (C) Indicated technique factors. The indicated technique factors shall be accurate to meet manufacturer's specifications. If these specifications are not available from the manufacturer, the factors shall be accurate to within plus or minus 10% of the indicated setting.

    (D) Tomographic plane indication and alignment.

      (i) For any single tomogram system, means shall be provided to permit visual determination of the tomographic plane or a reference plane offset from the tomographic plane.

      (ii) For any multiple slice tomogram system, means shall be provided to permit visual determination of the location of a reference plane. The reference plane can be offset from the location of the tomographic planes.

      (iii) If a device using a light source is used to satisfy the requirements of clause (i) or (ii) of this subparagraph, the light source shall provide illumination levels sufficient to permit visual determination of the location of the tomographic plane or reference plane under ambient light conditions of up to 500 lux.

    (E) Indication of CT conditions of operation. The CT system shall be designed such that the CT conditions of operation to be used during a scan or a scan sequence are indicated prior to the initiation of a scan or a scan sequence. On CT systems having all or some of these conditions of operation at fixed values, this requirement may be met by permanent markings. Indication of CT conditions of operation shall be visible from any position from which scan initiation is possible.

    (F) Initiation of operation.

      (i) The CT control panel and gantry shall provide visual indication whenever x rays are produced and, if applicable, whether the shutter is open or closed.

      (ii) Means shall be provided to require operator initiation of each individual scan or series of scans.

      (iii) All emergency buttons/switches shall be clearly labeled as to their functions.

    (G) Termination of exposure.

      (i) Means shall be provided to terminate the x-ray exposure automatically by either de-energizing the x-ray source or shuttering the x-ray beam in the event of equipment failure affecting data collection. Such termination shall occur within an interval that limits the total scan time to no more than 110% of its preset value through the use of either a backup timer or devices that monitor equipment function.

      (ii) A signal visible to the operator shall indicate when the x-ray exposure has been terminated through the means required by clause (i) of this subparagraph.

      (iii) The operator shall be able to terminate the x-ray exposure at any time during a scan or series of scans under CT system control, of greater than 0.5 seconds duration. Termination of the x-ray exposure shall necessitate resetting of the CT conditions of operation prior to initiation of another scan.

    (H) Additional requirements applicable to CT systems. Additional requirements applicable to CT systems containing a gantry manufactured after September 3, 1985, are as follows.

      (i) The total error in the indicated location of the tomographic plane or reference plane shall not exceed 5 mm.

      (ii) If the x-ray production period is less than 0.5 seconds, the indication of x-ray production shall be actuated for at least 0.5 seconds. Indicators at or near the gantry shall be discernible from any point external to the patient opening where insertion of any part of the human body into the primary beam is possible.

      (iii) The deviation of indicated scan increment versus actual increment shall not exceed plus or minus 1 mm with any mass from 0 to 100 kilograms (kg) resting on the support device. The patient support device shall be incremented from a typical starting position to the maximum incremented distance or 30 cm, whichever is less, and then returned to the starting position. Measurement of actual versus indicated scan increment can be taken anywhere along this travel.

  (2) Facility design requirements shall include the following.

    (A) Provision shall be made for two-way aural communication between the patient and the operator at the control panel.

    (B) Windows, mirrors, closed-circuit television, or an equivalent shall be provided to permit continuous observation of the patient during irradiation and shall be so located that the operator can observe the patient from the control panel.

      (i) Should the viewing system described in subparagraph (B) of this paragraph fail or be inoperative, treatment shall not be performed with the unit until the system is restored.

      (ii) In a facility that has a primary viewing system by electronic means and an alternate viewing system, should the viewing system described in subparagraph (B) of this paragraph fail or be inoperative, treatment shall not be performed with the unit until one of the systems is restored.

  (3) Measurements of the radiation output of the CT system, using the computed tomography dose index (CTDI) as recommended by the American Association of Physicists in Medicine (AAPM) and the International Council on Radiation Protection (ICRP), shall be performed by a licensed medical physicist.

    (A) Performance of the radiation output measurements shall be:

      (i) within 30 days after initial installation;

      (ii) thereafter annually or at intervals not to exceed 14 months from the date of the prior radiation output measurements;

      (iii) within 30 days of any major maintenance, that could affect radiation output; and

      (iv) within 30 days of any major change in equipment operation for example, introduction of a new software package.

    (B) The registrant shall ensure that radiation output measurements of a CT system are performed with a calibrated dosimetry system in accordance with subsection (i)(14) of this section.

  (4) A maintenance schedule shall be developed and followed. This schedule shall be in writing or documented in an electronic reporting system and shall be included in the registrant's operating and safety procedures. The maintenance schedule shall include but not be limited to the following:

    (A) radiation output measurements required by paragraph (3)(A) of this subsection;

    (B) acquisition of images by a licensed medical physicist obtained with phantoms and using the same processing mode and CT conditions of operation as are used to perform radiation output measurements required by paragraph (3)(A) of this subsection; and

    (C) acquisition of images by the registrant for quality control purposes obtained with phantoms and using protocol and intervals recommended by the manufacturer or the licensed medical physicist.

  (5) The registrant shall maintain the images specified in paragraph (4)(B) and (C) of this subsection in accordance with subsection (s)(1) of this section for inspection by the agency. The images may be maintained by either of the following methods:

    (A) photographic copies of the images obtained from the image display device; or