RULE §289.230Certification of Mammography Systems and Mammography Machines Used for Interventional Breast Radiography

    (B) Screen-film contact. Testing for screen-film contact shall be conducted using 40 mesh copper screen. The entire area of the cassette that may be clinically exposed shall be tested. This shall include all cassettes used for mammography in the facility.

    (C) Compression device performance. The maximum compression force for the initial power drive shall be between 25 pounds and 45 pounds. The system shall be capable of compressing the breast with a force of at least 25 pounds and shall be capable of maintaining this compression for at least 15 seconds.

  (5) Annual quality control tests. These tests shall be performed at an interval not to exceed (14) months.

    (A) Automatic exposure control performance. The AEC shall be capable of maintaining film optical density within plus or minus 0.15 of the mean optical density when thickness of a homogeneous material is varied over a range of 2 to 6 cm and the kVp is varied appropriately for such thicknesses over the kVp range and in the AEC mode used clinically in the facility.

    (B) Kilovoltage peak accuracy and reproducibility. At the most commonly used clinical settings of kVp, the coefficient of variation of reproducibility of the kVp shall be equal to or less than 0.02. The kVp shall be accurate to within plus or minus 5.0% of the indicated or selected kVp at the following:

      (i) the lowest clinical kVp that can be measured by a kVp test device;

      (ii) the most commonly used clinical kVp; and

      (iii) the highest available clinical kVp.

    (C) Focal spot condition. Facilities shall evaluate focal spot condition by determining the system resolution as follows.

      (i) Each system used for mammography, in combination with the mammography screen-film combination used in the facility, shall provide a minimum resolution of 11 cycles/millimeter (mm) (line-pairs/mm) when a high contrast resolution bar test pattern is oriented with the bars perpendicular to the anode-cathode axis, and a minimum resolution of 13 line-pairs/mm when the bars are parallel to that axis.

      (ii) The bar pattern shall be placed 4.5 cm above the breast support surface, centered with respect to the chest wall edge of the image receptor, and with the edge of the pattern within 1 cm of the chest wall edge of the image receptor.

      (iii) When more than one target material is provided, the measurement in clause (i) of this subparagraph shall be made using the appropriate focal spot for each target material.

      (iv) When more than one SID is provided, the test shall be performed at the SID most commonly used clinically.

      (v) Test kVp shall be set at the value used clinically by the facility for a standard breast and shall be performed in the AEC mode, if available. If necessary, a suitable absorber may be placed in the beam to increase exposure times. The screen-film cassette combination used by the facility shall be used to test for this requirement and shall be placed in the normal location used for clinical procedures.

    (D) Beam quality and half-value layer (HVL). The HVL shall meet the specifications of Title 21, CFR, §1020.30(m)(l) for the minimum HVL. These values, extrapolated to the mammographic range, are shown as follows. This test is performed using the clinical kVp on the standard breast. Values not shown in Table I may be determined by linear interpolation or extrapolation.

Attached Graphic

    (E) Breast entrance air kerma and AEC reproducibility. The coefficient of variation for both air kerma and mAs shall not exceed 0.05.

    (F) Dosimetry. The average glandular dose delivered during a single craniocaudal view of an FDA accepted phantom simulating a standard breast shall not exceed 3.0 milligray (mGy) (0.3 rad) per exposure.

    (G) X-ray field/light field/image receptor/compression paddle alignment. All systems shall meet the following.

      (i) All systems shall have beam-limiting devices that allow the entire chest wall edge of the x-ray field to extend to the chest wall edge of the image receptor and provide means to assure that the x-ray field does not extend beyond any edge of the image receptor by more than 2.0% of the SID.

      (ii) If a light field that passes through the x-ray beam limitation device is provided, it shall be aligned with the x-ray field so that the total of any misalignment of the edges of the light field and the x-ray field along either the length or the width of the visually defined field at the plane of the breast support surface shall not exceed 2.0% of the SID.

      (iii) The chest wall edge of the compression paddle shall not extend beyond the chest wall edge of the image receptor by more than 1.0% of the SID when tested with the compression paddle placed above the breast support surface at a distance equivalent to standard breast thickness. The shadow of the vertical edge of the compression paddle shall not be visible on the image.

    (H) Uniformity of screen speed. Uniformity of screen speed of all the cassettes in the facility shall be tested and the difference between the maximum and minimum optical densities shall not exceed 0.30. Screen artifacts shall also be evaluated during this test.

    (I) System artifacts. System artifacts shall be evaluated with a high-grade, defect-free sheet of homogeneous material large enough to cover the mammography cassette and shall be performed for all cassette sizes used in the facility using a grid appropriate for the cassette size being tested. System artifacts shall also be evaluated for all available focal spot sizes and target filter combinations used clinically.

    (J) Radiation output. The system shall be capable of producing a minimum output of 7.0 mGy air kerma per second (800 milliroentgen (mR) per second) when operating at 28 kVp in the standard mammography mode at any SID where the system is designed to operate. The system shall be capable of maintaining the required minimum radiation output averaged over a 3.0 second period.

    (K) Decompression. If the system is equipped with a provision for automatic decompression after completion of an exposure or interruption of power to the system, the system shall be tested to confirm that it provides the following:

      (i) an override capability to allow maintenance of compression;

      (ii) a continuous display of the override status; and

      (iii) a manual emergency compression release that can be activated in the event of power or automatic release failure.

    (L) The technique settings used for subparagraph (F) of this paragraph and paragraph (2) of this subsection shall be those used by the facility for its clinical images of a standard breast.

  (6) Densitometer and sensitometer. The calibration of the densitometer and sensitometer must be in accordance with the manufacturer's specifications.

  (7) Quality control tests - other modalities. For systems with image receptor modalities other than screen-film, the quality assurance program shall be substantially the same as the quality assurance program recommended by the image receptor manufacturer, except that the maximum allowable dose shall not exceed the maximum allowable dose for screen-film systems in paragraph (5)(F) of this subsection.

  (8) Mobile service operation. The registrant shall verify that mammography machines used to produce mammograms at more than one location meet the requirements in paragraphs (1) - (7) of this subsection. In addition, at each examination location, before any examinations are conducted, the registrant shall verify satisfactory performance of the mammography machines by using a test method that establishes the adequacy of the image quality produced by the machine. Processor performance shall be in accordance with paragraph (1) of this subsection.

  (9) Use of test results. After completion of the tests specified in paragraphs (1) - (8) of this subsection, the following shall occur.

    (A) The registrant shall compare the test results to the corresponding specified action limits; or, for nonscreen-film modalities, to the manufacturer's recommended action limits; or for post-move, pre-examination testing of mobile mammography machines, to the limits established in the test method used by the facility.

    (B) Components of the mammography system that fail quality assurance tests shall have corrective actions as indicated in the following.

      (i) If components in subclause (I) and (II) of this clause fail, corrective action shall be taken before any mammography films are processed:

        (I) paragraph (1) of this subsection describing processor quality control; and

        (II) paragraph (4)(A) of this subsection describing darkroom fog;

      (ii) If components in subclause (I) - (VI) of this clause fail, corrective action shall be taken before any mammography examinations are performed:

        (I) paragraph (2) of this subsection describing phantom image quality;

        (II) paragraph (4)(B) of this subsection describing screen-film contact;

Cont'd...