TITLE 25HEALTH SERVICES
PART 1DEPARTMENT OF STATE HEALTH SERVICES
CHAPTER 289RADIATION CONTROL
SUBCHAPTER EREGISTRATION REGULATIONS
RULE §289.232Radiation Control Regulations for Dental Radiation Machines

    (H) Radiation machines needing correction or repair. The correction or repair shall begin within 30 days following the failure and the registrant shall perform or cause to be performed the correction or repair according to a designated plan. Correction or repair shall be completed no longer than 90 days from discovery unless authorized in writing by the agency.

    (I) Records of radiation machine corrections or repairs. The registrant shall maintain records of corrections or repairs and any tests, measurements or numerical readings listed in subparagraph (J) of this paragraph in accordance with subsection (k)(2) of this section for inspection by the agency.

    (J) Equipment performance evaluations (EPE).

      (i) For all dental radiation machines, the registrant shall perform, or cause to be performed, EPE tests for each item specified in clauses (v) - (xi) of this subparagraph as follows:

        (I) within 30 days after initial installation of radiation machines:

        (II) within 30 days after reinstallation of a radiation machine; and

        (III) within 30 days after repair of a radiation machine component that would affect the radiation output that includes, but is not limited to, the timer, tube, and power supply.

      (ii) Frequency of EPE. For x-ray and CT systems, an EPE shall be performed at the frequency listed in the following table.

Attached Graphic

      (iii) Records of the EPE results shall be available for inspection by the agency and shall include the following:

        (I) measurements and numerical readings;

        (II) indication of pass or fail for each test; and

        (III) maintenance by the registrant in accordance with subsection (k)(2) of this section for inspection by the agency.

      (iv) Radiation machines needing correction or repair. If a radiation machine requires correction or repair following an EPE, the correction or repair shall begin within 30 days following the failure and the registrant shall perform or cause to be performed the correction or repair according to a designated plan. Correction or repair shall be completed no longer than 90 days from discovery unless authorized in writing by the agency.

      (v) Timer.

        (I) The accuracy of the timer shall meet the manufacturer's specifications. If the manufacturer's specifications are not obtainable, the timer accuracy shall be plus or minus 10% of the indicated time with testing performed at 0.5 second.

        (II) Means shall be provided to terminate the exposure at a preset time interval, a preset product of current and time, a preset number of pulses, or a preset radiation exposure to the image receptor. In addition, it shall not be possible to make an exposure when the timer is set to a "zero" or "off" position if either position is provided.

      (vi) Exposure reproducibility. When all technique factors are held constant, including control panel selections associated with automatic exposure control systems, the coefficient of variation of exposure for both manual and automatic exposure control systems shall not exceed 0.05. This requirement applies to clinically used techniques.

      (vii) Kilovolt peak. If the registrant possesses documentation of the appropriate manufacturer's kilovolt peak specifications, the radiation machine shall meet those specifications. If the registrant does not possess documentation of the appropriate manufacturer's kilovolt peak specifications, the indicated kilovolt peak shall be accurate to within plus or minus 10% of the indicated settings. For radiation machines with fewer than three fixed kilovolt peak settings, the radiation machine shall be checked at those settings.

      (viii) Tube stability. The x-ray tube shall remain physically stable during exposures. In cases where tubes are designed to move during exposure, the registrant shall assure proper and free movement of the radiation machine.

      (ix) Collimation. Field limitation shall meet the requirements of paragraphs (9) and (10) of this subsection.

      (x) Entrance exposure limits (air kerma limits) for dental facilities. The in-air exposure (entrance air kerma) for an adult bite wing view shall be determined from the exposure technique used by the registrant for the average adult patient. The in-air exposure (entrance air kerma) for intraoral (bite wing) dental radiography shall not exceed the following entrance exposure limits (air kerma limits):

Attached Graphic

      (xi) Measurements of the radiation output for a radiation machine. Measurements of the radiation output for a radiation machine shall be performed with a calibrated dosimetry system in accordance with the following.

        (I) The dosimetry system calibration shall be traceable to a national standard.

        (II) Dosimetry systems shall be calibrated within 24 months from the date of the prior calibration.

      (xii) Record of dosimetry system calibration. The registrant shall verify all dosimetry equipment meets the requirements of clause (xi) of this subparagraph.

  (6) Dental research.

    (A) Any research using radiation machines on humans shall be approved by an Investigational Review Board (IRB) as required by Title 45, CFR, Part 46, and Title 21, CFR, Part 56. The IRB shall include at least one licensed dentist to direct any use of radiation in accordance with this section.

    (B) Facilities with radiation machines with investigational device exemptions that are involved in clinical studies shall comply with primary regulations that govern the conduct of clinical studies and that apply to the manufacturers, sponsors, clinical investigators, institutional review boards, and the medical device. These regulations include the following:

      (i) 21 CFR, Part 812, Investigational Device Exemptions;

      (ii) 21 CFR, Part 50, Protection of Human Subjects;

      (iii) 21 CFR, Part 56, Institutional Review Boards;

      (iv) 21 CFR, Part 54, Financial Disclosure by Clinical Investigators; and

      (v) 21 CFR, Part 820, Subpart C, Design Controls of the Quality System Regulation.

  (7) Educational facilities. Facilities conducting training using non-humans are held to all the requirements of this section except for paragraph (5)(J) of this subsection concerning EPE and for paragraphs (12) and (13) of this subsection concerning image processing.

  (8) Certified radiation machines for dental facilities. The registrant shall not make, nor cause to be made, any modification of components or installations of components certified in accordance with the United States Food and Drug Administration Title 21, CFR, Part 1020, "Performance Standards for Ionizing Radiation Emitting Products," as amended, in any manner that could cause the installations or the components to fail to meet the requirements of the applicable parts of the standards specified in Title 21, CFR, Part 1020, except where a variance has been granted by the Director, Center for Devices and Radiological Health, United States Food and Drug Administration. A copy of the variance shall be maintained by the registrant in accordance with subsection (k)(2) of this section for inspection by the agency. All modifications of components or installation of components must be approved by the manufacturer.

  (9) Additional requirements for dental intraoral radiation machines.

    (A) Source-to-skin distance. Radiation machines designed for use with an intraoral image receptor shall be provided with means to limit source-to-skin distance to not less than:

      (i) 18 centimeters if operable above 50 kilovolt peak; or

      (ii) 10 centimeters if not operable above 50 kilovolt peak.

    (B) Field limitation. Radiation machines designed for use with an intraoral image receptor shall be provided with means to limit the x-ray beam such that:

      (i) if the minimum source-to-skin distance is 18 centimeters or more, the x-ray field at the minimum source-to-skin distance shall be restricted to a dimension of no more than seven centimeters; and

      (ii) if the minimum source-to-skin distance is less than 18 centimeters, the x-ray field at the minimum source-to-skin distance shall be restricted to a dimension of no more than six centimeters.

  (10) Additional requirements for dental extraoral radiation machines.

    (A) Dental panoramic radiation machines shall be provided with means to restrict the x-ray beam to the following:

      (i) the imaging slit in the transverse axis; and

      (ii) no more than a total of 0.5 inches larger than the imaging slit in the vertical axis.

    (B) All other dental extraoral radiation machines (e.g., cephalometric) shall be provided with means to restrict the x-ray field to the image receptor. The x-ray field shall not exceed the image receptor by more than:

Cont'd...