TITLE 25HEALTH SERVICES
PART 1DEPARTMENT OF STATE HEALTH SERVICES
CHAPTER 289RADIATION CONTROL
SUBCHAPTER EREGISTRATION REGULATIONS
RULE §289.232Radiation Control Regulations for Dental Radiation Machines

      (i) 2.0% of the source-to-image receptor distance for the length or width of the image receptor for rectangular collimation; or

      (ii) 2.0% of the source-to-image receptor distance for the diagonal of the image receptor for circular or polygon collimation.

  (11) Additional operational controls.

    (A) When a patient or image receptor must be held in position during radiography, mechanical supporting or restraining devices shall be used when the exam permits except in individual cases in which the registrant has determined that the holding devices are contraindicated.

    (B) The registrant's written operating and safety procedures required by paragraph (2) of this subsection shall include the following:

      (i) a list of circumstances in which mechanical holding devices cannot be routinely utilized; and

      (ii) a procedure used for selecting an individual to hold or support the patient or image receptor.

    (C) The operator position during the exposure shall be such that the operator's exposure is as low as reasonably achievable and the operator is a minimum of six feet from the useful beam or behind a protective barrier. The operator shall maintain verbal, aural, and visual contact with the patient.

  (12) Automatic and manual film processing for dental facilities and mobile dental services.

    (A) Films shall be developed in accordance with the time-temperature relationships recommended by the film manufacturer. The specified developer temperature for automatic processing and the time-temperature chart for manual processing shall be posted in the processing area. If the registrant determines an alternate time-temperature relationship is more appropriate for a specific facility, that time-temperature relationship shall be documented and posted.

    (B) Chemicals shall be replaced according to the chemical manufacturer or supplier's recommendations or at an interval not to exceed three months.

    (C) Darkroom light leak tests shall be performed at intervals not to exceed six months.

    (D) Lighting in the film processing/loading area shall be maintained with the filter, bulb wattage, and distances recommended by the film manufacturer for that film emulsion or with products that provide an equivalent level of protection against fogging.

    (E) Corrections or repairs of the light leaks or other deficiencies in subparagraphs (B) - (D) of this paragraph shall be initiated within 72 hours after discovery and completed no longer than 15 days from detection of the deficiency unless a longer time is authorized by the agency. Records of the corrections or repairs shall include the date and initials of the individual performing these functions and the registrant shall maintain the records in accordance with subsection (k)(2) of this section for inspection by the agency.

    (F) Documentation of the items in subparagraphs (B), (C), and (E) of this paragraph shall be maintained at the site where performed and shall include the date and initials of the individual completing these items. These records shall be made and maintained in accordance with subsection (k)(2) of this section for inspection by the agency.

  (13) Alternative processing systems. Users of daylight processing systems, laser processors, self-processing film systems, or other alternative processing systems shall follow manufacturer's recommendations for image processing. Documentation that the registrant is following manufacturer's recommendations shall include the date and initials of the individual completing the document and shall be made and maintained at the authorized use location where performed in accordance with subsection (k)(2) of this section for inspection by the agency.

  (14) Digital imaging acquisition systems.

    (A) Users of digital imaging acquisition systems shall follow quality assurance/quality control (QA/QC) protocol for digital imaging established by the manufacturer.

      (i) The registrant shall include the protocols established in paragraph (2) of this subsection in its operating and safety procedures.

      (ii) The registrant shall document the frequency at which the quality assurance/quality control protocol is performed. Documentation shall:

        (I) include the date and initials of the individual completing the document and the images acquired; and

        (II) be maintained and available at the authorized use location where performed in accordance with subsection (k)(2) of this section for inspection by the agency.

    (B) If a protocol cannot be established by the manufacturer, it shall be developed and implemented by the registrant.

      (i) The QA/QC protocol, as developed and implemented by the registrant, shall include image quality testing for, but not limited to, spatial resolution, noise, artifacts and contrast by using a commercially purchased testing tool or an inanimate object of at least three varying densities.

        (I) Images shall be acquired with each x-ray image receptor at an interval not to exceed three months.

        (II) Test images shall be compared to previous test images to assess degradation of image quality.

        (III) If a radiation machine or components of the digital imaging acquisition system require correction or repair following a quality test, the correction or repair shall begin within 30 days following the failure and the registrant shall perform or cause to be performed the correction or repair according to a designated plan. Correction or repair shall be completed no longer than 90 days from discovery unless authorized in writing by the agency.

      (ii) The registrant shall include the protocols established in paragraph (2) of this subsection in its operating and safety procedures.

      (iii) The registrant shall document the frequency at which the quality assurance/quality control protocol is performed. Documentation shall:

        (I) include the date and initials of the individual completing the document and the images acquired; and

        (II) be maintained and available at the authorized use location where performed in accordance with subsection (k)(2) of this section for inspection by the agency.

(k) Records and reports.

  (1) General provisions for records and reports.

    (A) Each registrant shall maintain records at each site, including sites authorized by the certificate of registration, conditions, and records sites for mobile services. The records shall include those specified in paragraph (2) of this subsection and shall be maintained at the time interval indicated for inspection by the agency. These records may be maintained in electronic format. These records shall be accessible to radiation machine operators during working hours.

    (B) All records required by this section shall be accurate and factual.

    (C) Each registrant shall use the SI units gray, sievert, and coulomb per kilogram, or the special units rad, rem, and roentgen, including multiples and subdivisions, and shall clearly indicate the units of all quantities on records required by this section.

    (D) The registrant shall make a clear distinction among the quantities entered on the records required by this section, such as, total effective dose equivalent, shallow dose equivalent, lens dose equivalent, and deep dose equivalent.

    (E) Each record required by this section shall be legible throughout the specified retention period.

    (F) The record shall be the original, a reproduced copy, or a microform, if the authorized personnel authenticate the copy or microform and that the microform is capable of producing a clear copy throughout the required retention period.

    (G) The record may also be stored in electronic format with the capability for producing legible, accurate, and complete records during the required retention period.

    (H) The registrant shall maintain adequate safeguards against tampering with and loss of records.

    (I) Copies of records required in subsections (i)(5)(I) and (J), (j)(5)(J), and (j)(12)(F) of this section and by certificate of registration condition that are relevant to operations at an additional authorized use location shall be maintained at that location in addition to the main site specified on a certificate of registration in accordance with subsection (k)(2) of this section.

    (J) Subject to the limitations provided in the Texas Public Information Act, Government Code, Chapter 552, all information and data collected, assembled, or maintained by the agency are public records open to inspection and copying during regular office hours.