<<Prev Rule

Texas Administrative Code

Next Rule>>
TITLE 25HEALTH SERVICES
PART 1DEPARTMENT OF STATE HEALTH SERVICES
CHAPTER 289RADIATION CONTROL
SUBCHAPTER EREGISTRATION REGULATIONS
RULE §289.233Radiation Control Regulations for Radiation Machines Used in Veterinary Medicine

        (I) The air kerma rate due to transmission through the primary protective barrier combined with radiation through the image intensifier shall be determined by measurements averaged over an area of 100 cm2 with no linear dimension greater than 20 cm.

        (II) If the source is below the tabletop, the measurement shall be made with the input surface of the fluoroscopic imaging assembly positioned 30 cm above the tabletop.

        (III) If the source is above the tabletop and the SID is variable, the measurement shall be made with the end of the beam-limiting device or spacer as close to the tabletop as it can be placed, provided that it shall not be closer than 30 cm.

        (IV) Movable grids and compression devices shall be removed from the useful beam during the measurement.

        (V) The attenuation block shall be positioned in the useful beam 10 cm from the point of measurement of entrance air kerma rate and between this point and the input surface of the fluoroscopic imaging assembly.

        (VI) The collimator shall be fully open when the measurement is made.

      (iii) X-ray field.

        (I) Compliance with subclauses (II) - (VII) of this clause shall be determined with the beam axis perpendicular to the plane of the image receptor.

        (II) Fluoroscopic radiation machine with a fixed SID and the capability of a visible area of no greater than 300 cm2 shall be provided with either stepless adjustment of the x-ray field or a means to further limit the x-ray field at the image receptor to 125 cm2 less. If the fluoroscopic radiation machine is provided with stepless adjustment, the minimum x-ray field size at the maximum SID shall be less than or equal to 5 cm by 5 cm at the image receptor.

        (III) Fluoroscopic radiation machines with a variable SID or a fixed SID with the capability of a visible area of greater than 300 cm2 shall be provided with stepless adjustment of the field size. The minimum x-ray field size at the maximum SID shall be less than or equal to 5 cm by 5 cm at the image receptor.

        (IV) Neither the length nor the width of the x-ray field in the plane of the image receptor shall exceed that of the visible area of the image receptor by more than 3.0 percent of the SID. The sum of the excess length and the excess width shall be no greater than 4.0 percent of the SID.

        (V) For rectangular x-ray fields used with circular image receptors, the error in alignment shall be determined along the length and width dimensions of the x-ray field that pass through the center of the visible area of the image receptor.

        (VI) For fluoroscopic radiation machines with only a manual mode of collimation, the x-ray field produced shall be limited to the area of the spot-film cassette at 16 inches above tabletop. Additionally, during fluoroscopy, the beam shall be restricted to the area of the input phosphor.

        (VII) Spot-film devices shall meet the following additional requirements.

          (-a-) Means shall be provided between the source and the animal for adjustment of the x-ray field size in the plane of the film to the size of that portion of the film that has been selected on the spot-film selector.

            (-1-) Such adjustment shall be automatically accomplished except when the x-ray field size in the plane of the film is smaller than that of the selected portion of the film.

            (-2-) The total misalignment of the edges of the x-ray field with the respective edges of the selected portion of the image receptor along the length or width dimensions of the x-ray field in the plane of the image receptor shall not exceed 3.0 percent of the SID when adjusted for full coverage of the selected portion of the image receptor.

            (-3-) The sum, without regard to sign of the misalignment along any two orthogonal dimensions, shall not exceed 4.0 percent of the SID.

          (-b-) The center of the x-ray field in the plane of the film shall be aligned with the center of the selected portion of the film to within 2.0 percent of the SID.

    (B) Activation of the fluoroscopic tube. X-ray production in the fluoroscopic mode shall be controlled by a device that requires continuous pressure by the fluoroscopist for the entire time of the exposure (continuous pressure type switch). When recording serial fluoroscopic images, the fluoroscopist shall be able to terminate the x-ray exposures at any time but means may be provided to permit completion of any single exposure of the series in process.

    (C) Control of scattered radiation.

      (i) Fluoroscopic configuration, including fluoroscopic table designs, shall not permit any portion of any individual's body, except the head, neck, and extremities, to be exposed to scattered radiation emanating from above or below the tabletop unless the radiation has passed through not less than a total of 0.25 mm lead equivalent material. The material may be drapes, self-supporting curtains, or viewing shields, in addition to any lead equivalency provided by a protective apron.

      (ii) Where sterile fields or special procedures prohibit the use of normal protective barriers or drapes, all of the following conditions shall be met.

        (I) All persons in the room where fluoroscopy is performed shall wear protective aprons that provide a shielding equivalent of 0.35 mm of lead.

        (II) The fluoroscopic field size shall be reduced to the absolute minimum required for the procedure being performed (area of clinical interest).

        (III) Operating and safety procedures shall reflect the above conditions, and fluoroscopy personnel shall exhibit awareness of situations requiring the use or nonuse of the protective drapes.

      (iii) For image-intensified fluoroscopic radiation machines with only a manual mode of collimation, the x-ray field produced shall be limited to the area of the spot-film cassette at 40cm above tabletop. Additionally, during fluoroscopy, the beam shall be restricted to the area of the input phosphor.

  (7) Additional requirements for CT radiation machines.

    (A) Warning Label. Each registrant shall ensure that each radiation machine is labeled in a conspicuous manner that cautions individuals that radiation is produced when it is energized. This label shall be affixed in a clearly visible location on the face of the control unit.

    (B) Indicated technique factors. The indicated technique factors shall be accurate to meet manufacturer's specifications. If these specifications are not available from the manufacturer, the factors shall be accurate to within plus or minus 10 percent of the indicated setting.

    (C) Tomographic plane indication and alignment.

      (i) For any single tomogram system, means shall be provided to permit visual determination of the tomographic plane or a reference plane offset from the tomographic plane.

      (ii) For any multiple slice tomogram system, means shall be provided to permit visual determination of the location of a reference plane. The reference plane can be offset from the location of the tomographic planes.

      (iii) If a device using a light source is used to satisfy the requirements of clause (i) or (ii) of this subparagraph, the light source shall provide illumination levels sufficient to permit visual determination of the location of the tomographic plane or reference plane under ambient light conditions of up to 500 lux.

    (D) Initiation of operation.

      (i) The x-ray control and gantry shall provide visual indication whenever x-rays are produced and, if applicable, whether the shutter is open or closed.

      (ii) Means shall be provided to require operator initiation of each individual scan or series of scans.

      (iii) All emergency buttons/switches shall be clearly labeled as to their functions.

    (E) Termination of exposure.

      (i) Means shall be provided to terminate the x-ray exposure automatically by either de-energizing the x-ray source or shuttering the x-ray beam in the event of radiation machine failure affecting data collection. Such termination shall occur within an interval that limits the total scan time to no more than 110 percent of its preset value through the use of either a backup timer or devices that monitor radiation machine function.

      (ii) A signal visible to the operator shall indicate when the x-ray exposure has been terminated through the means required by clause (i) of this subparagraph.

      (iii) The operator shall be able to terminate the x-ray exposure at any time during a scan or series of scans under CT radiation machines control of greater than 0.5 seconds duration. Termination of the x-ray exposure shall necessitate resetting of the CT conditions of operation before initiation of another scan.

  (8) Additional requirements for therapeutic radiation machines, simulators and electronic brachytherapy devices.

    (A) General requirements.

Cont'd...

Next Page Previous Page

Link to Texas Secretary of State Home Page | link to Texas Register home page | link to Texas Administrative Code home page | link to Open Meetings home page