<<Prev Rule

Texas Administrative Code

Next Rule>>
TITLE 25HEALTH SERVICES
PART 1DEPARTMENT OF STATE HEALTH SERVICES
CHAPTER 289RADIATION CONTROL
SUBCHAPTER EREGISTRATION REGULATIONS
RULE §289.227Use of Radiation Machines in the Healing Arts

  (13) Gonadal shielding. Gonadal shielding shall be used on patients when the gonads are in or within 5 cm of the useful beam. This requirement does not apply if the shielding will interfere with the diagnostic procedure. Gonadal shielding shall be of at least 0.5 mm lead equivalent material.

  (14) Measurements of the radiation output for x-ray, fluoroscopic, and CT systems. Measurements of the radiation output of x-ray, fluoroscopic, and CT systems shall be performed with a calibrated dosimetry system in accordance with the following.

    (A) The dosimetry system calibration shall be traceable to a national standard.

    (B) Dosimetry systems shall be calibrated within 24 months from the date of the prior calibration.

    (C) Measurements of radiation output shall be performed with a dosimetry system that has been calibrated within the 24 months preceding the date of the measurement.

    (D) Record of the dosimetry system calibration shall include:

      (i) manufacturer's name, model and serial number of each calibrated instrument;

      (ii) date of the calibration; and

      (iii) name of the individual recording the information.

    (E) The registrant shall record the dosimetry system calibration information specified in subparagraph (D)(i) - (iii) of this paragraph and maintain that record in accordance with subsection (s)(1) of this section for inspection by the agency.

(j) Radiographic entrance exposure (air kerma) limits. The in-air exposure (air kerma) determined for the technique used by the registrant for the specified average human adult patient thickness for medical radiography shall not exceed the entrance exposure (air kerma) limits in the following Table I.

Attached Graphic

(k) Machine requirements for general x-ray and fluoroscopic systems.

  (1) Warning label. The warning label will meet the requirements of §289.231(z) of this title.

  (2) Mechanical support of tube head. The tube housing assembly shall be adjusted to remain stable during an exposure unless tube housing movement is a designed function of the x-ray system.

  (3) Battery charge indicator. On battery-powered x-ray generators, visual means shall be provided on the control panel to indicate whether the battery is in a state of charge adequate for proper operation.

  (4) Beam quality. The following requirements apply to beam quality.

    (A) Half-value layer.

      (i) The half-value layer of the useful beam for a given x-ray tube potential shall not be less than the values shown in the following Table II. If it is necessary to determine such half-value layer at an x-ray tube potential that is not listed in Table II, linear interpolation may be made.

Attached Graphic

      (ii) For capacitor energy storage equipment, compliance with the requirements of this paragraph shall be determined with the maximum quantity of charge per exposure.

    (B) Filtration controls.

      (i) For x-ray systems that have variable kVp and variable filtration for the useful beam, a device shall link the kVp selector with the filters and shall prevent an exposure unless the minimum amount of filtration required by subparagraph (A) of this paragraph is in the useful beam for the given kVp that has been selected.

      (ii) Any other system having removable filters shall be required to have the minimum amount of filtration as required by subparagraph (A)(i) of this paragraph permanently located in the useful beam during each exposure.

  (5) Multiple tubes. Where two or more radiographic tubes are controlled by one exposure switch, the tube or tubes that have been selected shall be clearly indicated prior to initiation of the exposure. This indication shall be both on the x-ray control panel and at or near the tube housing assembly that has been selected.

  (6) Technique and exposure indicators.

    (A) The technique factors to be used during an exposure shall be indicated before the exposure begins except when automatic exposure controls are used, in which case the technique factors that are set prior to the exposure shall be indicated.

    (B) On x-ray systems having fixed technique factors, the requirement of subparagraph (A) of this paragraph may be met by permanent markings.

    (C) The x-ray control panel shall provide visual indication of the production of x-rays.

    (D) The indicated technique factors shall be accurate to meet manufacturer's specifications. If these specifications are not available from the manufacturer, the factors shall be accurate to within plus or minus 10% of the indicated setting.

  (7) X-ray control. An x-ray control shall be incorporated into each x-ray system such that an exposure can be terminated by the operator at any time except for an exposure of 0.5 seconds or less or during serial radiography when means shall be provided to permit completion of any single exposure of the series in process.

(l) Additional machine requirements for x-ray systems. This subsection does not apply to fluoroscopic or CT systems.

  (1) Beam limitation. Beam limitation shall be as follows.

    (A) Stationary general purpose x-ray systems.

      (i) Beam-limiting devices shall restrict the useful beam to the area of clinical interest as follows:

        (I) the misalignment of the x-ray field for a manual rectangular collimator shall be within plus or minus 2.0% of the SID for the length or width of the image receptor;

        (II) the x-ray field for a circular or polygon collimator shall not exceed the diagonal of the image receptor by more than 2.0% of the SID; or

        (III) the misalignment of the x-ray field for an automatic or semi-automatic collimator shall be within plus or minus 3.0% of the SID for the length and width of the image receptor and shall be within plus or minus 4.0% of the SID, without regard to the sign, of the sum of the difference of the length and width of the image receptor.

      (ii) If the area of clinical interest is smaller than the image receptor, the operator of the x-ray system shall use all beam limiting devices necessary to restrict the useful beam to the clinical area of interest.

      (iii) A method shall be provided for visually defining the perimeter of the x-ray field. The total misalignment of the edges of the visually defined field with the respective edges, either the length or width, of the x-ray field shall not exceed 2.0% of the SID.

      (iv) A numerical SID indicator shall be present and shall be accurate to within plus or minus 2.0% of the SID.

      (v) The system shall indicate when the axis of the x-ray field is perpendicular to the plane of the image receptor.

      (vi) The center of the x-ray field, when perpendicular to the image receptor, shall be accurate to within plus or minus 2.0% of the SID with respect to the center of the image receptor.

      (vii) The beam-limiting device shall numerically indicate the field size in the plane of the image receptor.

      (viii) Indication of field size dimensions and SIDs shall be specified in inches and/or centimeters.

      (ix) The field size indicated on the beam-limiting device shall be within plus or minus 2.0% of the SID along the width and length, separately, of the actual x-ray field size.

    (B) Portable x-ray systems. Portable x-ray systems shall comply with the requirements in subparagraph (A) of this paragraph, as applicable, based on manufacturer's design.

    (C) X-ray systems designed for one image receptor size. X-ray systems designed for only one image receptor size at a fixed SID shall provide a means to do the following:

      (i) limit the x-ray field to no greater than the dimensions of the image receptor at the SID, and to align the center of the x-ray field with the center of the image receptor to within plus or minus 2.0% of the SID center; or

      (ii) align the x-ray field such that the x-ray field does not extend beyond any edge of the image receptor at the SID.

    (D) Special purpose x-ray systems.

      (i) When the x-ray beam is perpendicular to the plane of the image receptor, a means shall be provided to do the following:

        (I) limit the x-ray field such that the x-ray field does not exceed each dimension of the image receptor by more than 2.0% of the SID; and

        (II) align the center of the x-ray field with the center of the image receptor to within plus or minus 2.0% of the SID.

Cont'd...

Next Page Previous Page

Link to Texas Secretary of State Home Page | link to Texas Register home page | link to Texas Administrative Code home page | link to Open Meetings home page