<<Prev Rule

Texas Administrative Code

Next Rule>>
TITLE 25HEALTH SERVICES
PART 1DEPARTMENT OF STATE HEALTH SERVICES
CHAPTER 289RADIATION CONTROL
SUBCHAPTER EREGISTRATION REGULATIONS
RULE §289.227Use of Radiation Machines in the Healing Arts

    (C) The x-ray control panel shall provide visual indication of the production of x-rays.

    (D) The indicated technique factors shall be accurate to meet manufacturer's specifications. If these specifications are not available from the manufacturer, the factors shall be accurate to within plus or minus 10% of the indicated setting.

  (7) X-ray control. An x-ray control shall be incorporated into each x-ray system such that an exposure can be terminated by the operator at any time except for an exposure of 0.5 seconds or less or during serial radiography when means shall be provided to permit completion of any single exposure of the series in process.

(l) Additional machine requirements for x-ray systems. This subsection does not apply to fluoroscopic or CT systems.

  (1) Beam limitation. Beam limitation shall be as follows.

    (A) Stationary general purpose x-ray systems.

      (i) Beam-limiting devices shall restrict the useful beam to the area of clinical interest as follows:

        (I) the misalignment of the x-ray field for a manual rectangular collimator shall be within plus or minus 2.0% of the SID for the length or width of the image receptor;

        (II) the x-ray field for a circular or polygon collimator shall not exceed the diagonal of the image receptor by more than 2.0% of the SID; or

        (III) the misalignment of the x-ray field for an automatic or semi-automatic collimator shall be within plus or minus 3.0% of the SID for the length and width of the image receptor and shall be within plus or minus 4.0% of the SID, without regard to the sign, of the sum of the difference of the length and width of the image receptor.

      (ii) If the area of clinical interest is smaller than the image receptor, the operator of the x-ray system shall use all beam limiting devices necessary to restrict the useful beam to the clinical area of interest.

      (iii) A method shall be provided for visually defining the perimeter of the x-ray field. The total misalignment of the edges of the visually defined field with the respective edges, either the length or width, of the x-ray field shall not exceed 2.0% of the SID.

      (iv) A numerical SID indicator shall be present and shall be accurate to within plus or minus 2.0% of the SID.

      (v) The system shall indicate when the axis of the x-ray field is perpendicular to the plane of the image receptor.

      (vi) The center of the x-ray field, when perpendicular to the image receptor, shall be accurate to within plus or minus 2.0% of the SID with respect to the center of the image receptor.

      (vii) The beam-limiting device shall numerically indicate the field size in the plane of the image receptor.

      (viii) Indication of field size dimensions and SIDs shall be specified in inches and/or centimeters.

      (ix) The field size indicated on the beam-limiting device shall be within plus or minus 2.0% of the SID along the width and length, separately, of the actual x-ray field size.

    (B) Portable x-ray systems. Portable x-ray systems shall comply with the requirements in subparagraph (A) of this paragraph, as applicable, based on manufacturer's design.

    (C) X-ray systems designed for one image receptor size. X-ray systems designed for only one image receptor size at a fixed SID shall provide a means to do the following:

      (i) limit the x-ray field to no greater than the dimensions of the image receptor at the SID, and to align the center of the x-ray field with the center of the image receptor to within plus or minus 2.0% of the SID center; or

      (ii) align the x-ray field such that the x-ray field does not extend beyond any edge of the image receptor at the SID.

    (D) Special purpose x-ray systems.

      (i) When the x-ray beam is perpendicular to the plane of the image receptor, a means shall be provided to do the following:

        (I) limit the x-ray field such that the x-ray field does not exceed each dimension of the image receptor by more than 2.0% of the SID; and

        (II) align the center of the x-ray field with the center of the image receptor to within plus or minus 2.0% of the SID.

      (ii) The requirements of clause (i) of this subparagraph may be met with a system that meets the requirements for a general purpose x-ray system as specified in subparagraph (A)(i) - (iv) of this paragraph or, when alignment means are also provided, may be met with either of the following:

        (I) an assortment of removable, fixed-aperture, beam-limiting devices sufficient to meet the requirement for each combination of image receptor size and SID for which the unit is designed with each such device having clear and permanent markings to indicate the image receptor size and SID for which it is designed; or

        (II) a beam-limiting device having multiple fixed apertures sufficient to meet the requirement for each combination of image receptor size and SID for which the radiation machine is designed. Permanent, clearly legible markings shall indicate the image receptor size and SID for which each aperture is designed and shall indicate which aperture is in position for use.

  (2) Radiation exposure control devices. Radiation exposure control devices shall include the following:

    (A) Timers. Means shall be provided to terminate the exposure at a preset time interval, a preset product of current and time, a preset number of pulses, or a preset radiation exposure to the image receptor. In addition, it shall not be possible to make an exposure when the timer is set to a "zero" or "off" position if either position is provided.

    (B) AEC. When AEC is provided, the following shall occur.

      (i) Indication shall be made on the control panel when AEC mode of operation is selected.

      (ii) If the x-ray tube potential is equal to or greater than 50 kVp, the minimum exposure time for field emission equipment rated for pulsed operation shall be equal to or less than a time interval equivalent to 2 pulses.

      (iii) The minimum exposure time for all x-ray systems other than that specified in clause (ii) of this subparagraph shall be equal to or less than 0.0167 second or a time interval required to deliver 5 mAs, whichever is greater.

      (iv) A visible and/or audible signal shall indicate when an exposure has been terminated.

    (C) Exposure interval reproducibility. When all technique factors are held constant, including control panel selections associated with AEC systems, the coefficient of variation of exposure interval for both manual and AEC systems shall not exceed 0.05. This requirement applies to clinically used techniques.

  (3) SSD. All portable x-ray systems shall be provided with means to limit the SSD to equal to or greater than 30 cm.

  (4) Exposure reproducibility. When all technique factors are held constant, including control panel selections associated with AEC systems, the coefficient of variation of exposure for both manual and AEC systems shall not exceed 0.05. This requirement applies to clinically used techniques.

  (5) Linearity. The average ratios of exposure (air kerma) mR/mGy to the indicated mAs product obtained at any 2 consecutive mA or mAs settings shall not differ by more than 0.10 times their sum, where X1 and X2 are the average mR/mAs (mGy/mAs) values obtained at each of 2 consecutive tube current settings:

Attached Graphic

  (6) Radiation from capacitor. Radiation from capacitor energy storage equipment in standby status. Radiation emitted from the x-ray tube when the exposure switch or timer is not activated shall not exceed a rate of 2 milliroentgens per hour (mR/hr) or 0.02 milligray per hour (mGy/hr) at 5 cm from any accessible surface of the diagnostic source assembly, with the beam-limiting device fully open.

  (7) X-ray systems needing correction or repair. In order to achieve compliance with this chapter, the correction or repair shall begin within 30 days following the failure and shall be performed according to a plan designated by the registrant. Correction or repair shall be completed no longer than 90 days from discovery unless authorized by the agency.

  (8) Records of an x-ray system corrections or repairs. The registrant shall maintain records of corrections or repairs and any tests, measurements or numerical readings listed in subsection (o)(6) of this section in accordance with subsection (s)(1) of this section for inspection by the agency.

(m) Fluoroscopic systems and spot-film devices for all facilities.

  (1) Limitation of the useful beam. Limitation of the useful beam shall be as follows.

    (A) Primary barrier.

Cont'd...

Next Page Previous Page

Link to Texas Secretary of State Home Page | link to Texas Register home page | link to Texas Administrative Code home page | link to Open Meetings home page